| Model Number |
98814-65 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/19/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Customer reported receiving erratic readings on their adc blood glucose meter.Customer reported receiving readings of 327 mg/dl, 497 mg/dl and 151 mg/dl within 10 minutes.The results when plotted on a parkes error grid fell into the "c" zone showing the difference in values to be clinically significant.There was no report of death, serious injury, or mistreatment associated with this event.
|
| |
|
Additional Manufacturer Narrative
|
|
No product has been returned, extended investigation is pending at this time.A follow- up report will be submitted once additional information is obtained.The device manufacturer date for the reported meter is unknown.The date entered is the date abbott diabetes care became aware of the event.All pertinent information available to abbott diabetes care has been submitted.
|
| |
|
Event or Problem Description
|
|
Customer reported receiving erratic readings on their adc blood glucose meter.Customer reported receiving readings of 327 mg/dl, 497 mg/dl and 151 mg/dl within 10 minutes.The results when plotted on a parkes error grid fell into the "c" zone showing the difference in values to be clinically significant.There was no report of death, serious injury, or mistreatment associated with this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Repeated attempts by adc to retrieve the product were unsuccessful and/or the customer discarded the product.No product has been returned.To date, product has not been returned for further investigation.The useful life of exceed meter is 4 years.As the manufacturing date of this product is before 2009, it has been in distribution beyond its useful life at the time of the complaint.Since the product exceeded its useful life, it is determined to have met specification when the product was released and through its lifespan.Therefore, no further investigation activities are required for meter xce031-3323.Dhr for the precision strips were reviewed and the dhr showed the precision strips passed all tests before release.Retain testing was performed for the precision strips and all units performed within specification.If the product is returned, a physical investigation will be performed and a follow-up report submitted.All pertinent information available to abbott diabetes care has been submitted.
|
| |
|
Event or Problem Description
|
|
Customer reported receiving erratic readings on their adc blood glucose meter.Customer reported receiving readings of 327 mg/dl, 497 mg/dl and 151 mg/dl within 10 minutes.The results when plotted on a parkes error grid fell into the "c" zone showing the difference in values to be clinically significant.There was no report of death, serious injury, or mistreatment associated with this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Meter (b)(4) has been returned and investigated.Visual inspection has been performed on the returned meter and no issues were observed.Meter powered on with button depression and with strip insertion.Control solution testing was performed and no issues were observed.All results were within range specification.No malfunction or product deficiency was identified.If the reported test strips are returned, an investigation will be performed.An extended investigation has been performed for the reported complaint and there was no indication that the product did not meet specification.To date, product has not been returned for further investigation.The useful life of xceed meter is 4 years.As the manufacturing date of this product is before 2009, it has been in distribution beyond its useful life at the time of the complaint.Since the product exceeded its useful life, it is determined to have met specification when the product was released and through its lifespan.Therefore, no further investigation activities are required for meter xce031-3323.Dhr for the precision strips were reviewed and the dhr showed the precision strips passed all tests before release.Retain testing was performed for the precision strips and all units performed within specification.If the partial product is returned, a physical investigation will be performed and a follow-up report submitted.All pertinent information available to abbott diabetes care has been submitted.
|
| |
|
Search Alerts/Recalls
|