| Model Number |
120-260 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Seroma (2069)
|
| Date of Event |
10/14/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
|
| |
|
Event or Problem Description
|
|
Patient reported right side rupture.Device remains implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: analysis of the returned device identified: underweight and broken shell.A visual and microscopic analysis was performed which identified: striated broken on anterior.Based on the device analysis the final assessment is: a striated broken on anterior side assessed as surgical damage.
|
| |
|
Event or Problem Description
|
|
Physician reported "replacement due to rupture in right side" confirmed via ultrasound."bilateral breast pain and tighten sensation for a month" was reported via clinical chart."rt periprosthetic fluid collection," "r/o prosthetics injury," and "breast pain" were also reported via sonogram.The device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.
|
| |
|
Event or Problem Description
|
|
Physician reported "replacement due to rupture in right side" confirmed via ultrasound."bilateral breast pain and tighten sensation for a month" was reported via clinical chart."rt periprosthetic fluid collection," "r/o prosthetics injury," and "breast pain" were also reported via sonogram.The device has been explanted.
|
| |
|