• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALLERGAN (COSTA RICA) STYLE 120 SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALLERGAN (COSTA RICA) STYLE 120 SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED Back to Search Results
Model Number 120-260
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Codes Failure of Implant (1924); Seroma (2069)
Date of Event 10/14/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
 
Event or Problem Description
Patient reported right side rupture.Device remains implanted.
 
Additional Manufacturer Narrative
Device evaluation: analysis of the returned device identified: underweight and broken shell.A visual and microscopic analysis was performed which identified: striated broken on anterior.Based on the device analysis the final assessment is: a striated broken on anterior side assessed as surgical damage.
 
Event or Problem Description
Physician reported "replacement due to rupture in right side" confirmed via ultrasound."bilateral breast pain and tighten sensation for a month" was reported via clinical chart."rt periprosthetic fluid collection," "r/o prosthetics injury," and "breast pain" were also reported via sonogram.The device has been explanted.
 
Additional Manufacturer Narrative
A review of the device history record has been completed.No deviations or non-conformances noted.
 
Event or Problem Description
Physician reported "replacement due to rupture in right side" confirmed via ultrasound."bilateral breast pain and tighten sensation for a month" was reported via clinical chart."rt periprosthetic fluid collection," "r/o prosthetics injury," and "breast pain" were also reported via sonogram.The device has been explanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STYLE 120 SILICONE GEL FILLED BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
Manufacturer (Section G)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
Manufacturer Contact
chris sprague
12331-a riata trace parkway
building 3
austin, TX 78727
8479375615
MDR Report Key13239920
Report Number9617229-2022-00005
Device Sequence Number17281839
Product Code FTR
UDI-Device Identifier10888628002524
UDI-Public10888628002524
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
P020056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 06/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/26/2021
Device Model Number120-260
Device Catalogue Number120-260
Device Lot Number2942100
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 12/28/2021
Supplement Date Received by Manufacturer04/27/2022
06/06/2022
Initial Report FDA Received Date01/12/2022
Supplement Report FDA Received Date05/24/2022
06/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/26/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age43 YR
Patient SexFemale
Patient Weight54 KG
-
-