ENDOLOGIX AFX2; BIFURCATED STENT GRAFT Back to Search Results
Model Number BEA28-100/I16-40
Medical Device Problem Codes Adverse Event Without Identified Device or Use Problem (2993); Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 01/06/2022
Type of Reportable Event Serious Injury
Event or Problem Description
The patient was initially treated for an abdominal aortic aneurysm (aaa) with implant of an afx2 bifurcated stent graft and an afx vela suprarenal.Reportedly, the procedure went well; however, the aortic neck was challenging and there was an existing small dissection in the lcia (left common iliac artery).Final angiogram identified an indeterminate endoleak.The physician has elected to monitor the patient and do a 30 day or 60 follow-up ct (computed tomography) scan.The patient was reported to be discharged one day post procedure.
 
Additional Manufacturer Narrative
The devices involved in this event will not be returned for evaluation and remain implanted in the patient.Patient medical records and imaging studies will be requested for further evaluation by a clinical specialist.If additional information pertinent to the incident is obtained, a follow-up report will be submitted.Device iteration is afx2.
 
Additional Manufacturer Narrative
The reported adverse event/incident was investigated in alignment with endologix operating procedures and work instructions.Where possible, it is endologix practice to make at least three good faith efforts to retrieve a reported adverse event/incident related device as well as medical records and medical imaging.An evaluation of the manufacturing record was completed.A review of the part number/lot number combination needed for device identification shows the device to have been properly manufactured and released in accordance with the device master record.An evaluation of the device could not be completed.The device was not returned to endologix for evaluation because it remains implanted.A clinical evaluation of the reported adverse event/incident was completed.An examination of medical records and/or medical imaging received by endologix shows the indeterminate endoleak was a type ii endoleak (lumbar) unresolved on the 30-day post-index ct scan.This is not consistent with the reported adverse event/incident.Type ii endoleaks are anatomy-related events and are not caused or contributed by the device.Therefore, a complaint is no longer warranted since there is no alleged deficiency related to the identity, quality, durability, reliability, safety, effectiveness, or performance of the device.Device iteration is afx2.
 
Event or Problem Description
The patient was initially treated for an abdominal aortic aneurysm (aaa) with implant of an afx2 bifurcated stent graft and an afx vela suprarenal.Reportedly, the procedure went well; however, the aortic neck was challenging and there was an existing small dissection in the lcia (left common iliac artery).Final angiogram identified an indeterminate endoleak.The physician has elected to monitor the patient and do a 30 day or 60 follow-up ct (computed tomography) scan.The patient was reported to be discharged one day post procedure.After the initial report, additional information was received reporting that the one (1) month follow-up computerized tomography (ct) scan showed no endoleak.Subsequently, clinical analysis determined that the indeterminate endoleak was a type ii endoleak (lumbar) unresolved on the 30-day post-index ct scan.Type ii endoleaks are anatomy-related events and are not caused or contributed by the device.A complaint is not warranted since there is no alleged deficiency related to the identity, quality, durability, reliability, safety, effectiveness, or performance of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AFX2
Common Device Name
BIFURCATED STENT GRAFT
Manufacturer (Section D)
ENDOLOGIX
2 musick
irvine CA 92618
Manufacturer (Section G)
ENDOLOGIX
2 musick
,
irvine CA 92618
Manufacturer Contact
gary kirchgater
2 musick
,
irvine, CA 92618
8009832284
MDR Report Key13240041
Report Number2031527-2022-00006
Device Sequence Number4484186
Product Code MIH
UDI-Device Identifier00818009014450
UDI-Public(01)00818009014450(17)240519
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
P040002
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,Company Representative
Initial Reporter Occupation Physician
Remedial Action Other
Type of Report Initial,Followup
Report Date (Section B) 01/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/19/2024
Device Model NumberBEA28-100/I16-40
Device Lot Number2460707004
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 01/06/2022
Supplement Date Received by Manufacturer03/29/2022
Initial Report FDA Received Date01/12/2022
Supplement Report FDA Received Date04/05/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/11/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
AFX VELA SUPRARENAL (LN 2428431005)
Outcome Attributed to Adverse Event Other;
Patient Age83 YR
Patient SexMale
-
-