• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NUVASIVE SPECIALIZED ORTHOPEDICS, INC. MAGEC SPINAL BRACING AND DISTRACTION SYSTEM; GROWING ROD SYSTEM- MAGNETIC ACTUATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NUVASIVE SPECIALIZED ORTHOPEDICS, INC. MAGEC SPINAL BRACING AND DISTRACTION SYSTEM; GROWING ROD SYSTEM- MAGNETIC ACTUATION Back to Search Results
Model Number MS1-4590S
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 11/22/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device has not been returned for evaluation as it remains in-situ.The root cause is unable to be determined at this time.If any additional information is provided, a supplemental report will be submitted.
 
Event or Problem Description
Information was received that the rod has a broken pin and a broken radial bearing.It was reported the rod is still able lengthen.No patient adverse event was reported.
 
Additional Manufacturer Narrative
Additional data a2, a3, b3, b6, d6, g3, h2, h10.A review of the dhr documents indicated that the magec rod was manufactured by the specified requirements and met all of the required quality inspections before shipment.
 
Additional Manufacturer Narrative
Per the complaint description, the product will not be returned as it remains in-situ.Reportedly, the rod is currently distracting as expected, thus treatment is continuing.The x-ray image revealed a broken ball bearing and locking pin, which confirms the reported failure.A rod with a broken pin cannot transfer torque from the magnet module to the lead screw, which means no distraction is possible.This device can be distracted by pulling force only, in this case the pulling force would be coming from the pair rod, which is still fully functional.Because the locking pin is broken, there is nothing constraining the axial motion of the distraction rod within the housing tube.With the spine growing and the other rod lengthening, the broken rod can passively extend as the distraction rod slides within the housing tube.The pair rod is then able to drive the construct of the affected rod.The potential cause of the broken radial bearing and locking pin is excessive axial loading due to the patient's unique anatomy and/or daily activities.The iqc inspection data for the lots of radial ball bearing and locking pin were reviewed and the parts met design specifications per the engineering drawings confirming there were not any manufacturing or assembly defects.
 
Event or Problem Description
No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAGEC SPINAL BRACING AND DISTRACTION SYSTEM
Common Device Name
GROWING ROD SYSTEM- MAGNETIC ACTUATION
Manufacturer (Section D)
NUVASIVE SPECIALIZED ORTHOPEDICS, INC.
101 enterprise dr, ste 100
aliso viejo CA 92656
Manufacturer (Section G)
NUVASIVE SPECIALIZED ORTHOPEDICS, INC.
101 enterprise dr, ste 100
aliso viejo CA 92656
Manufacturer Contact
geoff gannon
101 enterprise dr, ste 100
aliso viejo, CA 92656
8583448112
MDR Report Key13240621
Report Number3006179046-2022-00110
Device Sequence Number11911686
Product Code PGN
UDI-Device Identifier00812258026295
UDI-Public812258026295
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
K161751
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMS1-4590S
Device Catalogue NumberMS1-4590S
Device Lot NumberA161012-04
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/20/2021
Supplement Date Received by Manufacturer01/19/2022
03/17/2022
Initial Report FDA Received Date01/12/2022
Supplement Report FDA Received Date02/04/2022
03/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/12/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age14 YR
Patient SexMale
-
-