SMITH & NEPHEW, INC. GII UNIV PS FEMORAL IMPACTOR; PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL
|
Back to Search Results |
|
| Model Number |
71440895 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/29/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that, during a total knee arthroplasty, a genesis ii universal femoral impactor broke inside the patient.No pieces fell into the patient.Surgery was resumed, without any delay, using a smith & nephew back-up device.Patient was not injured beyond the reported problem.
|
| |
|
Additional Manufacturer Narrative
|
|
The associated device was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The screw of the device is broken, rendering the device inoperative.The device shows signs of extensive use.A review of complaint history revealed similar events for the listed device, but no similar events for the batch, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.Assessment of historical escalated cases concluded that there are no prior actions related to this device and failure mode.A contribution of the device to the reported event could be corroborated as the device shows signs of damage/wear.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor future complaints and investigate as necessary.We consider this investigation closed.
|
| |
|
Search Alerts/Recalls
|
|
|