| Model Number |
N/A |
| Medical Device Problem Code |
Separation Failure (2547)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/17/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Concomitant medical products: part #: cm-9300, st 2.9mm gl drill, lot# 71241-1.Part #: cm-9129, quattro link anchor 2.9mm, lot#: 67061-2.Part #: cm-9129, quattro link anchor 2.9mm, lot#: 70603-2.The device will not be returned for analysis, as the device was discarded; however, an investigation of the reported event is in progress.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported during surgery on the right side of the patient, the device came apart leaving the center rod stuck on the implant.Surgeon used a needle driver and mallet to remove and the implant came out on its own.No adverse events have been reported as a result of the malfunction.Attempts have been made and additional information on the reported event is unavailable at this time.
|
| |
|
Additional Manufacturer Narrative
|
|
Upon investigation completion, it is now determined to be not reportable.No serious injury or harm to the patient reported for this event or prior events with same or similar products.The initial report should be voided.
|
| |
|
Event or Problem Description
|
|
Upon investigation completion, it is now determined to be not reportable.No serious injury or harm to the patient reported for this event or prior events with same or similar products.The initial report should be voided.
|
| |
|