| Model Number |
SOFTECHD+21.5 |
| Medical Device Problem Codes |
Improper or Incorrect Procedure or Method (2017); Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
12/13/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Based on the review of the batch documentation conducted, lenstec can confirm that all procedures in the manufacturing and packaging of the devices were conducted correctly.Additionally, we have not received any other complaints from these batches.
|
| |
|
Event or Problem Description
|
|
Lenstec received an email stating " broken haptic was torn off.Another lens successfully implanted.".
|
| |
|
Additional Manufacturer Narrative
|
|
Based on the review of the batch documentation conducted, lenstec can confirm that all procedures in the manufacturing and packaging of the devices were conducted correctly.Additionally, we have not received any other complaints from these batches.Furthermore, after device evaluation it was determined that the lens was handled whilst in a dehyrdrated state, indicating the manufactuers instructions for use were not followed.
|
| |
|
Event or Problem Description
|
|
Lenstec received an email stating " broken haptic was torn off.Another lens successfully implanted.".
|
| |
|