• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CARDIOVASCULAR SYSTEMS, INC. STEALTH ORBITAL ATHERECTOMY SYSTEM; PERIPHERAL ATHERECTOMY DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CARDIOVASCULAR SYSTEMS, INC. STEALTH ORBITAL ATHERECTOMY SYSTEM; PERIPHERAL ATHERECTOMY DEVICE Back to Search Results
Model Number PRD-SC30-MICRO
Medical Device Problem Code Entrapment of Device (1212)
Health Effect - Clinical Code Device Embedded In Tissue or Plaque (3165)
Date of Event 12/11/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The results of the investigation are inconclusive since the reported device was not returned for analysis.Based on the information received, the cause of the reported event could not be conclusively determined.Follow-up attempts to the site regarding the event details and patient information have been made, but no additional information has been received.If additional information is received a supplemental report will be submitted.Csi id: (b)(4).
 
Event or Problem Description
The diamondback coronary orbital atherectomy device (oad) was operated in the anterior tibial artery without issue.The oad was then advanced to the peroneal artery and became stuck in the vessel.The oad could not be removed and the patient was sent to surgery.The patient was stable following the procedure.
 
Event or Problem Description
The diamondback coronary orbital atherectomy device (oad) was operated in the anterior tibial artery without issue.The oad was advanced to the peroneal artery, through a total chronic occlusion and while spinning became stuck in the vessel.Troubleshooting attempts including inflating a balloon near the oad, attaching a hemostat to the catheter and spinning counterclockwise and inserting a guide catheter over the csi device were made, but were unsuccessful.The oad could not be removed and the patient was sent to surgery.The device was successfully removed and the patient was stable following the procedure.
 
Additional Manufacturer Narrative
The device history record for this oad lot number has been reviewed.No issues or discrepancies were noted during this review that would have contributed to the reported event.The device met material, assembly, and quality control requirements.Csi id: 11148.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEALTH ORBITAL ATHERECTOMY SYSTEM
Common Device Name
PERIPHERAL ATHERECTOMY DEVICE
Manufacturer (Section D)
CARDIOVASCULAR SYSTEMS, INC.
1225 old highway 8 nw
saint paul MN 55112
Manufacturer (Section G)
CARDIOVASCULAR SYSTEMS, INC.
1225 old highway 8 nw
saint paul MN 55112
Manufacturer Contact
brittany leider
1225 old highway 8 nw
saint paul, MN 55112
6512591600
MDR Report Key13252043
Report Number3004742232-2022-00009
Device Sequence Number19940543
Product Code MCW
UDI-Device Identifier10850000491264
UDI-Public(01)10850000491264(17)230630(10)385178-1
Combination Product (Y/N)N
Initial Reporter StateGU
Initial Reporter CountryUS
PMA/510(K) Number
K190634
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2023
Device Model NumberPRD-SC30-MICRO
Device Catalogue Number7-10059-01
Device Lot Number385178-1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Fifteen-Day
Initial Date Received by Manufacturer 12/18/2021
Supplement Date Received by Manufacturer01/18/2022
Initial Report FDA Received Date01/13/2022
Supplement Report FDA Received Date02/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/13/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-