Model Number 1012447-12
Medical Device Problem Codes Material Rupture (1546); Improper or Incorrect Procedure or Method (2017)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/05/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the procedure was to treat an unspecified lesion with heavy calcification and mild tortuosity.A 2.50 x 12mm nc trek rx balloon dilatation catheter (bdc) was advanced to the target lesion without issues.However, during the third inflation at 22 atmospheres, the balloon ruptured.The bdc was removed from the patient.It is unknown at this time how the target lesion was treated.There were no adverse patient effects and there was no clinically significant delay in the procedure.No additional information was provided.
 
Additional Manufacturer Narrative
The device is not returning for evaluation.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event or Problem Description
Subsequent to the initially filed report, the following information was added: it was reported that the procedure was to treat a lesion in the left anterior descending artery (lad) with heavy calcification and mild tortuosity.The bdc was removed from the patient and the procedure was completed with a non-abbott balloon.No additional information was provided.
 
Additional Manufacturer Narrative
The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history of the reported lot did not indicate a lot specific quality issue.The investigation determined the reported balloon rupture appear to be related to user error/operational context.It was reported t device was inflated three times at 22 atmospheres (atms) each.On the third inflation at 22 atms, the balloon ruptured.It should be noted that the coronary dilatation catheters (cdc), nc trek rx, global, instruction for use (ifu) states: balloon pressure should not exceed the rated burst pressure (rbp).The rbp for the nc trek rx device is 18 atm.In this case, it is likely the combination of the ifu deviation and the heavily calcified anatomy resulted in the reported balloon rupture.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.H6: device code 2017 improper or incorrect procedure or method -above rbp.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NC TREK CORONARY DILATATION CATHETER
Common Device Name
CORONARY DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR COSTA RICA, REG # 3009564766
52 calle 3 b31 coyol free zone
el coyol alajuela
CS  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key13252544
Report Number2024168-2022-00473
Device Sequence Number4584818
Product Code LOX
UDI-Device Identifier08717648151750
UDI-Public08717648151750
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K103153
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2022
Device Model Number1012447-12
Device Catalogue Number1012447-12
Device Lot Number91127G1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 12/24/2021
Supplement Date Received by Manufacturer01/14/2022
Initial Report FDA Received Date01/13/2022
Supplement Report FDA Received Date01/25/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/27/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-