Model Number DB-1140-S
Medical Device Problem Code Battery Problem (2885)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/30/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient's implantable pulse generator (ipg) was depleted.The physician has requested an analysis to determine if the battery depletion was normal.
 
Additional Manufacturer Narrative
Correction to the initial mdr in block b5.Analysis on the returned ipg revealed that the ipg could be linked with a remote control but could not be linked with an interrogative software tool.Attempts to bring up the device from the end of service state using an external power source failed.The printed circuit board level electrical test revealed normal device characteristics.The ipg was no longer responding thus the longevity of the device could not be determined.The device was damaged during the explant procedure due to an unknown source; it could not be confirmed if electrocautery was used during the procedure.The complaint of ipg depletion after 22 months could not be investigated as the device was no longer functioning and the data log was not retrievable.The cause of the complaint could not be established.
 
Event or Problem Description
It was reported that the patient's primary cell implantable pulse generator (ipg) had reached the end of life battery message.The patient underwent a routine battery replacement procedure and was doing fine post-operatively.The physician requested an analysis to determine if the battery depletion was normal.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERCISE PC
Common Device Name
STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN SYMPTOMS
Manufacturer (Section D)
BOSTON SCIENTIFIC NEUROMODULATION
25155 rye canyon loop
valencia CA 91355
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
talar tahmasian
25155 rye canyon loop
valencia, CA 91355
6619494863
MDR Report Key13252744
Report Number3006630150-2022-00045
Device Sequence Number19981835
Product Code NHL
UDI-Device Identifier08714729984467
UDI-Public08714729984467
Combination Product (Y/N)N
Initial Reporter StateID
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date12/13/2020
Device Model NumberDB-1140-S
Device Catalogue NumberDB-1140-S
Device Lot Number631181
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/30/2021
Supplement Date Received by Manufacturer04/13/2022
Initial Report FDA Received Date01/13/2022
Supplement Report FDA Received Date05/04/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/13/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age76 YR
Patient SexMale
-
-