| Model Number |
SC-3138-35 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/04/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that during the permanent implant procedure, the lead extension was observed to be damaged at the sheathing; the sheathing skidded down the wires from the connector.The patient underwent an explant of the lead extension.
|
| |
|
Event or Problem Description
|
|
It was reported that during the permanent implant procedure, the lead extension was observed to be damaged at the sheathing; the sheathing skidded down the wires from the connector.The patient underwent an explant of the lead extension.
|
| |
|
Additional Manufacturer Narrative
|
|
Device analysis of returned lead extension sc-3138-35 (b)(6) included a visual inspection that revealed the lead was cleanly cut near the connector of the lead and the proximal end portions of the lead was not returned.The clean-cut damage is a result of a typical explant procedure, and it is not considered a failure.No other anomalies were identified on the returned portion of the lead.
|
| |
|