|
WRIGHTS LANE SYNTHES USA PRODUCTS LLC TFNA FENESTRATED HELICAL BLADE 105MM; APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT
|
Back to Search Results |
|
| Model Number |
04.038.405 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Nonunion/Delayed-union Bone Fracture (2369)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Additional narrative: without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that on an unknown date, a nonunion from a previous surgery in the femoral shaft/proximal femur was discovered.The surgeon removed a tfna device, used the ria to collect bone graft, and then inserted a 95 degree blade plate to correct the nonunion.No further information was provided.This report is for one (1) tfna fenestrated helical blade 105mm.This is report 2 of 5 for (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H6: the device lot number is unknown, therefore a device history record (dhr) review could not be performed.If more information become available, the record will be re-assessed.Investigation summary: the complaint device was not received for investigation.A photo investigation was performed based on the received photo.The images were reviewed, and the complaint condition is not confirmed.After review of the x-rays provided, it is shown that a set of one tfna nail, one tfna blade, 2 locking screws screws and one end cap are present implanted on the patient in the femoral shaft/proximal femur area.The implants do not appear to have any defect or malfunction that could contribute to the nonunion condition or the adverse events reported by the patient.No other issues were found on the device.This complaint is not confirmed as a photo inspection does not show malfunction of the complaint device.A manufacturing record evaluation cannot be performed due to lack of lot number.A definite assignable root cause could not be determined based on the provided information.As the device was not returned, and as-received condition could not be assessed and a dimensional inspection and document/specification review were not completed.No dhr review was completed since no lot number was supplied via photo or additional information.During the investigation no product design issues or discrepancies were observed (based on the images) that may have contributed to the complaint condition.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|
|
|