Model Number 9733560
Medical Device Problem Code Communication or Transmission Problem (2896)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/20/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No patient information provided as no patient was involved in this concern.Concomitant medical products: other relevant device(s) are: product id: 960-152, version #: 3.17.The software investigation found that the reported event was related to a software issue.This issue was documented in a medtronic navigation software anomaly tracking database.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received information regarding the navigation system.It was reported that some korean domestic cone beam ct (pointnix, hdx) are not compatible with medtronic navigation somehow, especially fusion and fusion compact, so that demonstration was not able to proceed.The system showed 'no image loaded'.Even though the manufacturer representative (rep) checked it contains one x-ray image, the rep was not able to find the main reason why it does not load in the old system.The issue has not been resolved yet.Additional information received from a manufacturer representative indicated no further actions would be taken.It was confirmed the site that the navigation system in question was not compatible because of the program setting (media io).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FUSION® ENT NAVIGATION SYSTEM
Common Device Name
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key13254157
Report Number1723170-2022-00058
Device Sequence Number14598561
Product Code HAW
UDI-Device Identifier00763000235079
UDI-Public00763000235079
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K001284
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 01/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number9733560
Device Catalogue Number9733560
Device Lot NumberN08203999
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/06/2021
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/05/2022
Initial Report FDA Received Date01/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/16/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-