| Model Number |
391235 |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/23/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
An orsiro drug-eluting stent system was selected for treatment of a pre-dilated lesion.The stent dislodged inside the guiding catheter.The guiding catheter with the stent inside was removed.The intervention was completed with another orsiro stent.
|
| |
|
Additional Manufacturer Narrative
|
|
Neither the complaint instrument nor the angiographic material was returned for analysis.Therefore, no technical investigation on the subject could be performed.The product release documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for the reported event.Review of the product release documentation for the product detailed above verified that the instrument was manufactured according to specifications and fulfilled all the requirements of in-process and final inspection.During final inspection, every stent system undergoes visual inspection to ensure correct stent embedding and homogeneous crimping of the stent.Further, the stent outer diameter is verified to 100 percent and the stent retention force of a defined number of samples is tested from every lot.Based on the conducted investigations of the device being subject to this complaint, no material or manufacturing related root cause was determined.
|
| |
|
Search Alerts/Recalls
|