| Model Number |
A34-34/C100-O20 V |
| Medical Device Problem Codes |
Off-Label Use (1494); Unintended Movement (3026); Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Codes |
Aneurysm (1708); Failure of Implant (1924)
|
| Date of Event |
05/25/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The patient was initially implanted with an afx2 bifurcated stent graft, two (2) vela suprarenal stent grafts and a limb stent graft to treat an abdominal aortic aneurysm (aaa).Approximately four (4) years post initial implant during a routine follow up, the patient was identified with a type 3a endoleak with component separation.The physician elected to implant another suprarenal stent graft and an infrarenal stent graft to treat this event.The patient was reported as stable post re-intervention.This event was previously reported under mdr # 2031527-2021-00056.Approximately three (3) months post re-intervention, an unknown proximal endoleak, an unknown distal type 3 endoleak (type unknown) and aneurysm sac growth were identified.No further information was provided, and the patient status was not reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The device involved in this event will not be returned for evaluation as it remains implanted in the patient.Patient medical records and imaging studies will be requested for further evaluation by the clinical specialist.If additional information pertinent to the incident is obtained, a follow-up report will be submitted.Device iteration is afx2.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported adverse event/incident was investigated in alignment with endologix operating procedures and work instructions.Where possible, it is endologix practice to make at least three good faith efforts to retrieve a reported adverse event/incident related device as well as medical records and medical imaging.An evaluation of the manufacturing record was completed.A review of the part number/lot number combination needed for device identification shows the device to have been properly manufactured and released in accordance with the device master record.A clinical evaluation of the adverse event/incident was completed.An examination of medical records and/or medical imaging received by endologix shows that the indeterminate endoleak and sac growth of 11.3mm complaints are confirmed.This is consistent with the reported adverse event/incident.Procedure related harms, device, user, procedure or anatomy relatedness of this complaint could not be determined with the lack of medical records and/or imaging available for review.Additionally, the clinical evaluation also shows that there was evidence to reasonably suggest stent movement of a proximal extension of 8mm occurred that was not included in the event as reported and there was a non endologix limb placed in the right external iliac artery.It is unclear if this contributed to the reported event.No additional investigation of this reported adverse event/incident is planned.However, should additional information relevant to the investigation outcome become available, a follow-up report will be submitted.Endologix will continue to monitor this and similar adverse events/incidents.Device iteration is afx2.Corrections: b5: describe event or problem -updated.D1: brand name has been updated.D2: product description has been updated.D4: model number has been updated.D4: catalog number has been updated.D4: lot # has been updated.D4: expiration date has been updated.D4: unique identifier (udi) # has been updated.D10: therapy dates has been updated.G3: date received by manufacturer - updated.H4: device manufacture date has been updated.H6: medical device problem codes: remove code 3190.H6: investigation finding codes: remove code 3233.H6: investigation conclusion codes: remove code 11.
|
| |
|
Event or Problem Description
|
|
The patient was initially implanted with an afx2 bifurcated stent graft, two (2) vela suprarenal stent grafts and a limb stent graft to treat an abdominal aortic aneurysm (aaa).Approximately four (4) years post initial implant during a routine follow up, the patient was identified with a type 3a endoleak with component separation.The physician elected to implant another suprarenal stent graft and an infrarenal stent graft to treat this event.The patient was reported as stable post re-intervention.This event was previously reported under mdr # 2031527-2021-00056.Approximately three (3) months post re-intervention, an unknown proximal endoleak, an unknown distal type 3 endoleak (type unknown) and aneurysm sac growth were identified.No further information was provided, and the patient status was not reported.Additional information: clinical evaluation shows that a non endologix limb was placed in the right external iliac artery.Clinical evaluation also shows at there was evidence to reasonably suggest stent movement of a proximal extension of 8mm occurred that was not included in the event as reported.
|
| |
|