MAUDE Adverse Event Report: ZOLL CIRCULATION ZOLL IVTM COOL LINE CATHETER; CENTRAL VENOUS CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL CIRCULATION ZOLL IVTM COOL LINE CATHETER; CENTRAL VENOUS CATHETER Back to Search Results
Model Number CL-2295
Medical Device Problem Codes Fluid/Blood Leak (1250); Material Integrity Problem (2978)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/22/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Zoll has not received the product for investigation.A follow-up report will be submitted when the product is returned and investigation has been completed.
 
Event or Problem Description
During ivtm therapy, customer noticed blood tinge in the start-up kit (suk) tubing connecting the thermogard ivtm system to the cool line catheter.The patient was not reaching the target temperature at 36 degrees, patient temperature stayed at 36.5 degrees.The cool line catheter (lot #unknown) suspected to be leaking.The customer placed the thermogard ivtm system on standby and per reporter, the console is functioning properly.The catheter in question was removed from the patient subclavian vein and ice packs was used to continue with treatment.Per reporter, when slip tip syringe was applied to orange luer and aspirated, blood was noted in the syringe indicates a potential catheter leak.No other lines were placed near the cool line catheter.No consequences or impact to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZOLL IVTM COOL LINE CATHETER
Common Device Name
CENTRAL VENOUS CATHETER
Manufacturer (Section D)
ZOLL CIRCULATION
2000 ringwood ave.
san jose CA 95131
Manufacturer (Section G)
ZOLL CIRCULATION
2000 ringwood ave.
san jose CA 95131
Manufacturer Contact
kim thoa nguyen
2000 ringwood ave.
san jose, CA 95131
4084192922
MDR Report Key13256421
Report Number3010617000-2022-00042
Device Sequence Number5023446
Product Code NCX
UDI-Device Identifier00849111075046
UDI-Public00849111075046
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K101987
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 01/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCL-2295
Device Catalogue Number8700-0654-01
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/22/2021
Initial Report FDA Received Date01/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-