• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET 3I 3I T3® EX HEX TAPERED IMPLANT 4 X 8.5MM; DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET 3I 3I T3® EX HEX TAPERED IMPLANT 4 X 8.5MM; DENTAL IMPLANT Back to Search Results
Model Number BOET485
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/20/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Zimmer biomet complaint number (b)(4).
 
Event or Problem Description
It was reported that the patient came with a fallen prosthesis and a broken prosthesis holding screw, the implant is also fractured.Implant removed.
 
Event or Problem Description
No further event information is available at the time of this report.
 
Additional Manufacturer Narrative
Zimmerbiomet complaint number (b)(4).The following sections have been updated: b4: date of this report.B5: describe event or problem.D9: device available for evaluation change ¿no' to 'yes'.G3: date received by the manufacturer.H1: type of report, follow-up number.H2: follow up type.H3: device evaluated by manufacturer: change ¿no' to 'yes'.H6: evaluation codes.H10: additional narrative.An implant and unknown biomet screw were returned for investigation.Visual inspection of the as returned products identified worn markings due to usage, excessive bone/debris and a fracture on the threads on both implant and screw (screw inside implant).No pre-existing conditions were noted on the per.The reported device was located on tooth #26 and was used for approximately 1 year.Pictures or x-ray images were not provided.Review of appropriate documentation: documents reviewed: biomet 3i dental implant ifu (p-iis086gi) rev h - july 2021 information identified: 'warnings' 'precautions' 'potential adverse events'.Documents reviewed: biomet 3i restorative products ifu (p-iis086gr) rev f ¿ october 2019 information identified: 'warnings' 'precautions' 'potential adverse events'.Dhr review: dhr review was completed for the subject lot number (2011080851).It was confirmed that all operations and inspections were executed as per applicable procedure.No deviations or non-conformances, which could have caused or contributed to the reported event was noted as part of the dhr.Lot was inspected and passed all acceptance criteria by qa.Dhr review could not be performed, as the lot number associated with the reported unknown biomet screw product is not available.Zimmer biomet quality management system (qms) has controls in place to ensure the distribution of conforming product within specifications.Complaint history review: complaint history review was performed for the reported lot number (2011080851) for similar event and no other complaint was identified.No complaint history review could be performed without relevant lot and item information unknown biomet screw.Review completed utilizing keywords: fracture : implant and fracture : screw.Post market trend review: april post market trending was reviewed and there were no actionable events or corrective actions for the reported event or device.Based on the available information, device malfunction did occur and the reported event was confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3I T3® EX HEX TAPERED IMPLANT 4 X 8.5MM
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
susanne taylor
4555 riverside drive
palm beach gardens, FL 33410
5617766700
MDR Report Key13257037
Report Number0001038806-2022-00047
Device Sequence Number19218977
Product Code DZE
UDI-Device Identifier00844868033295
UDI-Public(01)00844868033295(17)240613(10)2019060194
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K133049
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date (Section B) 06/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date06/13/2024
Device Model NumberBOET485
Device Catalogue NumberBOET485
Device Lot Number2019060194
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/22/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/22/2021
Supplement Date Received by Manufacturer06/28/2022
Initial Report FDA Received Date01/14/2022
Supplement Report FDA Received Date06/30/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/13/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
Patient Weight96 KG
-
-