• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION MAVERICK; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION MAVERICK; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS Back to Search Results
Model Number 7580
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/20/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that balloon rupture occurred.The patient presented with coronary artery disease (cad).The de novo target lesion was located in the coronary vessel.A 2.00mm x 12mm maverick balloon catheter was advanced for dilatation.However, it was found out that the balloon ruptured.The procedure was completed with another maverick balloon catheter.There were no complications nor injuries reported and the patient was in stable condition.
 
Additional Manufacturer Narrative
(a2) age at time of event: 18 years or older.Device evaluated by mfr: returned product consisted of a maverick 2mr balloon catheter.The device was visually and microscopically examined.There were numerous kinks to the hypotube of the device.There was both contrast and blood in the inflation lumen and balloon.There was blood in the guidewire lumen and the balloon was loosely folded.The device was prepped with an inflation device filled with water and connected to the inflation port to inflate the device.There was a pinhole in the balloon located 16mm from the tip of the device.The device failed to inflate to rbp.Product analysis confirmed the reported event, as during functional testing the device was found to have a pinhole damage to the balloon.
 
Event or Problem Description
It was reported that balloon rupture occurred.The patient presented with coronary artery disease (cad).The de novo target lesion was located in the coronary vessel.A 2.00mm x 12mm maverick balloon catheter was advanced for dilatation.However, it was found out that the balloon ruptured.The procedure was completed with another maverick balloon catheter.There were no complications nor injuries reported and the patient was in stable condition.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAVERICK
Common Device Name
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
pmt 741 persiaran cassia selat
bandarcassia, pulau pinan 14110
MY   14110
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key13257117
Report Number2134265-2022-00248
Device Sequence Number11140094
Product Code LOX
UDI-Device Identifier08714729369967
UDI-Public08714729369967
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
P860019
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/12/2024
Device Model Number7580
Device Catalogue Number7580
Device Lot Number0026617952
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/29/2021
Supplement Date Received by Manufacturer03/25/2022
Initial Report FDA Received Date01/14/2022
Supplement Report FDA Received Date04/01/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/12/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-