STRYKER TRAUMA KIEL K-WIRE GAMMA Ø3,2X450 MM; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Model Number 1210-6450
Medical Device Problem Code Mechanical Jam (2983)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/17/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that cutter got stuck on a third party attachment.The third party technician stated that the cutter might have contributed to the malfunction.
 
Additional Manufacturer Narrative
The device will not be returned.If additional information becomes available, it will be provided in a supplemental report.Device disposition is unknown.
 
Additional Manufacturer Narrative
The reported event was confirmed on provided image.The stryker k-wire was stuck in a non-stryker adapter.A review of the device history could not be performed as the lot code could not be identified.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.The ifu states: ¿ stryker surgical instruments are intended to be used in surgical procedures in conjunction with stryker implants which come with package inserts of their own.¿ thus, the item was not used according to the instruction; the event was caused by off-label use.In case substantive information or the item itself becomes available we reserve the right to reopen the investigation and to reassess the root cause.
 
Event or Problem Description
It was reported that cutter got stuck on a third party attachment.The third party technician stated that the cutter might have contributed to the malfunction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
K-WIRE GAMMA Ø3,2X450 MM
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM  D-24232
Manufacturer (Section G)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM   D-24232
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key13257205
Report Number0009610622-2022-00016
Device Sequence Number14642748
Product Code LXH
UDI-Device Identifier04546540175359
UDI-Public04546540175359
Combination Product (Y/N)N
Initial Reporter CountrySZ
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1210-6450
Device Catalogue Number12106450
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/17/2021
Supplement Date Received by Manufacturer04/28/2022
Initial Report FDA Received Date01/14/2022
Supplement Report FDA Received Date05/26/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-