• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS WINTER & IBE GMBH JAWS "HIQ+", 5 X 330 MM, GRASPING FORCEPS, CROCEOLMI; HAND INSTRUMENTS, STANDARD, BIPOLAR AND MONOPOLAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS WINTER & IBE GMBH JAWS "HIQ+", 5 X 330 MM, GRASPING FORCEPS, CROCEOLMI; HAND INSTRUMENTS, STANDARD, BIPOLAR AND MONOPOLAR Back to Search Results
Model Number WA64150A
Medical Device Problem Codes Break (1069); Fracture (1260); Material Fragmentation (1261); Mechanical Problem (1384)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/06/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The suspect medical device has not yet been returned to olympus for evaluation/investigation.Therefore, the exact cause of the user's experience and the reported phenomenon could not be determined and is being judged as unknown.However, if the suspect medical device is returned for evaluation/investigation or additional significant information becomes available, this report will be updated.
 
Event or Problem Description
Olympus was informed that during a therapeutic laparoscopic low anterior resection procedure the jaws at the distal end of the grasping forceps broke and a fragment fell into the patient.However, no fragment remained inside the patient since it was reportedly retrieved.The intended procedure was successfully completed with a similar device and there was no report about an adverse event or patient injury.
 
Additional Manufacturer Narrative
The suspect medical device was returned to the manufacturer for evaluation/investigation.The evaluation/investigation confirmed that one jaw at the distal end of the grasping forceps is broken off at the thinnest section.In view of the instrument¿s age, this damage is considered a fatigue fracture caused by wear and tear due to frequent use/reprocessing.There are numerous small superficial scratches on the instrument indicating frequent use of the instrument as well as discolorations on the jaws indicating frequent/inadequate reprocessing.Therefore, this event/incident was attributed to user error.Also, a material or quality problem can be excluded since a manufacturing and quality control review was performed for the affected lot number of the grasping forceps without showing any abnormalities.The case will be closed on olympus side with no further actions.The reported event/incident will be recorded for trending and surveillance purposes and the user will be informed about the investigation results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JAWS "HIQ+", 5 X 330 MM, GRASPING FORCEPS, CROCEOLMI
Common Device Name
HAND INSTRUMENTS, STANDARD, BIPOLAR AND MONOPOLAR
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg
Manufacturer Contact
daniel wladow
kuehnstrasse 61
hamburg 22045
GM   22045
4940669662955
MDR Report Key13257409
Report Number9610773-2022-00054
Device Sequence Number13818886
Product Code GCJ
UDI-Device Identifier04042761068970
UDI-Public04042761068970
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K944201
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility,Company Representative,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 03/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberWA64150A
Device Catalogue NumberWA64150A
Device Lot Number124W
Was Device Available for Evaluation? Device Returned to Manufacturer
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/06/2022
Supplement Date Received by Manufacturer03/23/2022
Initial Report FDA Received Date01/14/2022
Supplement Report FDA Received Date03/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/01/2012
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-