| Model Number |
Y399H |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/20/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.The following information was requested, but unavailable: what is the lot number? no further information is available.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent a cesarean section surgery on (b)(6) 2021 and suture was used.During the procedure, the needle detached from the suture while suturing the myometrium.No adverse patient consequences were reported.Additional information was requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Date sent to the fda: 02/24/2022.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H3 analysis: visual analysis of the returned sample revealed that one opened sample, of product code y399h, was received for evaluation.After investigation of the sample, short insertion was observed in the strand.The visual inspection concluded that the combination of the needle/suture was found detached since the insertion end of the suture did not meet the requirement.The defect is caused because the suture was not properly inserted inside of the needle barrel hole resulting in a pull-out defect.Based on the information currently available, the pull-out was identified during the investigation of the sample received.This product issue will be addressed through ethicon inc¿s quality system.
|
| |
|
Search Alerts/Recalls
|