• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. PDS PLUS VIO 18IN 4-0 S/A PS-2 PRM MP; SUTURE, SURGICAL, ABSORBABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. PDS PLUS VIO 18IN 4-0 S/A PS-2 PRM MP; SUTURE, SURGICAL, ABSORBABLE Back to Search Results
Model Number PDP513G
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/22/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).This is a combination product, and the event has been reviewed for both the suture and the triclosan.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.Trade name - irgacare®, active ingredient(s) ¿ triclosan, dosage form ¿ suture/solid/parenteral, strength ¿ = 2360 ug/m.Events reported in 2210968-2022-00405.
 
Event or Problem Description
It was reported that a patient underwent an unknown surgery on (b)(6) 2021 and suture was used.During the procedure, the suture broke during interrupted suturing.No adverse patient consequences reported.No additional information was provided.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Date sent to the fda: (b)(6) 2022.Component code: g07002 no device problem found.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Investigation summary: three packets of product code pdp513g were returned for analysis with the packaging closed.Upon initial inspection of the samples, no external damages were observed on the packets.In order to evaluate the conditions of the returned samples, the packets were opened, and no defects were detected.The swage and attachment area was noted to be as expected.The sutures were dispensed without problems and examined along the strand to detect any issue related to damaged or breakage sutures and no defects were observed during evaluation.Functional test was performed, and the tensile strength result was above the minimum requirements.As part of our quality process, the manufacturing records of this lot-serial number were reviewed, and the manufacturing standards were met prior to the release of this batch.As part of quality process, all devices are manufactured, inspected, and released to approved specifications.There may have been other circumstances or issues that occurred during the use of the device that we were unable to duplicate during our laboratory analysis.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PDS PLUS VIO 18IN 4-0 S/A PS-2 PRM MP
Common Device Name
SUTURE, SURGICAL, ABSORBABLE
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-JUAREZ
avenida de las torres 7125
col salvacar
cuidad juarez
MX  
Manufacturer Contact
elba bello
1000 route 202
raritan, NJ 08869
MDR Report Key13257809
Report Number2210968-2022-00406
Device Sequence Number4760747
Product Code NEW
UDI-Device Identifier10705031048485
UDI-Public10705031048485
Combination Product (Y/N)Y
Initial Reporter CountryJA
PMA/510(K) Number
K061037
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2023
Device Model NumberPDP513G
Device Catalogue NumberPDP513G
Device Lot NumberRJMHLX
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/28/2022
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/21/2021
Supplement Date Received by Manufacturer01/28/2022
Initial Report FDA Received Date01/14/2022
Supplement Report FDA Received Date02/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/31/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-