• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH 1.5MM VA-LCKNG SCREW SLF-TPNG WITH T4 STARDRIVE RECESS 8MM; SCREW, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH 1.5MM VA-LCKNG SCREW SLF-TPNG WITH T4 STARDRIVE RECESS 8MM; SCREW, FIXATION, BONE Back to Search Results
Model Number 02.130.208
Medical Device Problem Code Material Integrity Problem (2978)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/03/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A definite assignable root cause could not be determined based on the provided information.As the device was not returned, and as-received condition could not be assessed and a dimensional inspection and document/specification review were not completed.During the investigation no product design issues or discrepancies were observed (based on the images) that may have contributed to the complaint condition.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Device history lot: part: 02.130.208, lot: 2l96949, manufacturing site: (b)(4), supplier: synthes gmbh, release to warehouse date: 21 january 2019, expiration date: n/a.A manufacturing record evaluation was performed for the finished article lot and no non-conformances were identified.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Device report from synthes reports an event in (b)(6) as follows: it was reported that on (b)(6), 2021, during an unknown hand procedure, the stardrive screws were discovered as damaged.No surgical delay or patient consequence reported.This report is for one (1) 1.5mm va-lckng screw slf-tpng with t4 stardrive recess 8mm this is report 1 of 4 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10: additional narrative: d9, h3, h6: the device was received, the investigation could not be completed, and no conclusion could be drawn, as product is entering the complaint system.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: h3, h6: the product was returned to us customer quality (cq) for evaluation.The us cq team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that the va lockscr ø1.5 self-tap l8 had a stripped stardrive.No other issues were identified.The dimensional inspection was not performed due to the post manufacturing damage.The observed condition of the va lockscr ø1.5 self-tap l8 was consistent with a random component failure that may have been caused by exposure to excessive/unintended forces.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the va lockscr ø1.5 self-tap l8 was observed to have a stripped stardrive.While no definitive root cause could be determined from the available information, it is probable that the va lockscr ø1.5 self-tap l8 was stripped due to the exposure to excessive/unintended forces.There was no indication that a design or manufacturing issue contributed to the complaint.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
1.5MM VA-LCKNG SCREW SLF-TPNG WITH T4 STARDRIVE RECESS 8MM
Common Device Name
SCREW, FIXATION, BONE
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
WERK GRENCHEN (CH)
solothurnstrasse 186
grenchen 2540
SZ   2540
Manufacturer Contact
kara ditty-bovard
1302 wrights lane east
west chester, PA 19380
6103142063
MDR Report Key13257904
Report Number8030965-2022-00339
Device Sequence Number14634434
Product Code HWC
UDI-Device Identifier10886982044624
UDI-Public(01)10886982044624
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
K141527
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number02.130.208
Device Catalogue Number02.130.208
Device Lot Number2L96949
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/22/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 12/21/2021
Supplement Date Received by Manufacturer02/22/2022
05/09/2022
Initial Report FDA Received Date01/14/2022
Supplement Report FDA Received Date03/01/2022
05/10/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/21/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
VA LOCKSCR Ø1.5 SELF-TAP L12 SST; VA LOCKSCR Ø1.5 SELF-TAP L16 SST; VA LOCKSCR Ø1.5 SELF-TAP L8
Patient SexUnknown
-
-