| Model Number |
00711811 |
| Medical Device Problem Codes |
Retraction Problem (1536); Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/25/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device subject of the event will be returned to steris endoscopy for evaluation.The device history record was reviewed and confirmed the device lot was manufactured to specification.Statements in the instructions for use include: "carefully remove the tip protector from the distal tip of the needle.Actuate the proximal luer back and forth, visually observing the needle being adequately advanced from the device's distal end.The spring-loaded handle assists in needle retraction; however, always visually check the retraction as well.Note: do not stretch the device because stretching may cause the needle projection length to be altered.Do not attempt to actuate the injection needle with the catheter in a straightened position.The injection needle has been engineered and manufactured with a specific "preload" that is necessary to assure the full projection and full retraction of the needle from the distal hub when the device is in very tortuous configurations." steris endoscopy has offered in-service training on the use of the carr-lock injection needle to the user facility; however, the user facility has declined.A follow-up mdr will be submitted when additional information becomes available.No additional issues have been reported.
|
| |
|
Event or Problem Description
|
|
The user facility reported via health (b)(6) report to steris that their carr-locke injection needle failed to retract.There was no report of harm to the patient or user of the device.
|
| |
|
Additional Manufacturer Narrative
|
|
The steris endoscopy engineering team evaluated the device and could not duplicate the reported event.The device was found to be operating according to specification.No additional issues have been reported.
|
| |
|
Search Alerts/Recalls
|