• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOTRONIK SE & CO. KG ITREVIA 7 DR-T DF-1; ICD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOTRONIK SE & CO. KG ITREVIA 7 DR-T DF-1; ICD Back to Search Results
Model Number 392412
Medical Device Problem Code Premature Elective Replacement Indicator (1483)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/05/2021
Type of Reportable Event Malfunction
Event or Problem Description
This device was explanted due to eri with unexpected battery behavior.No adverse patient events were reported.Should additional information be received, this file will be updated.
 
Additional Manufacturer Narrative
The device is currently not available for analysis.No conclusion can be drawn at this time.No additional information is available at the moment.The file is closed.The investigation will be re-opened should additional data become available.
 
Additional Manufacturer Narrative
Prior to the analysis of the device, the quality documents accompanying the manufacturing process for this icd were re-investigated.All production steps were performed accordingly, and in particular the final acceptance test proved the device functions to be as specified.Next, the device was interrogated.The interrogation could be properly performed and revealed the eri battery status, confirming the clinical observation.Next, the icd was subjected to an analysis of the electrical parameters.The current consumption of the icd was checked and found to be normal and as expected.Analysis of the battery condition, however, showed an inconsistency of charge taken from the battery and the battery voltage.A premature battery depletion could be confirmed during analysis.Please note, this icd is affected by the field safety corrective action, bio-lqc, initiated in march 2021.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ITREVIA 7 DR-T DF-1
Common Device Name
ICD
Manufacturer (Section D)
BIOTRONIK SE & CO. KG
woermannkehre 1
berlin 12359
Manufacturer Contact
6024 jean road
lake oswego, OR 97035
8772459800
MDR Report Key13258474
Report Number1028232-2022-00261
Device Sequence Number14618445
Product Code MRM
UDI-Device Identifier04035479128807
UDI-Public04035479128807
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P050023
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2018
Device Model Number392412
Device Catalogue NumberSEE MODEL NO.
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/13/2022
Supplement Date Received by Manufacturer01/18/2022
04/29/2022
Initial Report FDA Received Date01/14/2022
Supplement Report FDA Received Date01/19/2022
05/03/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/04/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
Patient Age53 YR
Patient SexMale
-
-