• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARDEX® URETHRAL CATHETER ROBINSON MODEL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARDEX® URETHRAL CATHETER ROBINSON MODEL Back to Search Results
Model Number 056110
Medical Device Problem Codes Flaked (1246); Device Contaminated During Manufacture or Shipping (2969)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the robinson intermittent catheter packaging was flaked off onto other components inside the pack.The provided lots were (ngdz3241), (ngez0776), (ngfn2700).
 
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event or Problem Description
It was reported that the robinson intermittent catheter packaging was flaked off onto other components inside the pack.The provided lots were (ngdz3241), (ngez0776), (ngfn2700).
 
Additional Manufacturer Narrative
The reported event is confirmed manufacturing related.1 sample was confirmed to exhibit the reported failure.The product had caused the reported failure.Visual evaluation of the attached photo sample noted 1 unopened (with original packaging), ureteral catheter.Visual inspection of the sample noted flacking of the strip packaging inside the sealed packaging on the catheter.This is out of specification, which states "loose foreign matter, engraved and spots >1/32¿ is not allowed".A potential root cause for this could be not following the production areas cleaning procedure.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.A labeling review was not performed because labeling could not have prevented the reported failure.H11: section a through f - the information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDEX® URETHRAL CATHETER ROBINSON MODEL
Common Device Name
URETHRAL CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
juan velez
8195 industrial blvd
covington 30014
7707846100
MDR Report Key13259729
Report Number1018233-2022-00035
Device Sequence Number16569266
Product Code KOD
UDI-Device Identifier00801741033506
UDI-Public(01)00801741033506
Combination Product (Y/N)N
Initial Reporter StateLA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date01/31/2025
Device Model Number056110
Device Catalogue Number056110
Device Lot NumberNGFN2700
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/01/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 01/01/2022
Supplement Date Received by Manufacturer08/10/2022
Initial Report FDA Received Date01/14/2022
Supplement Report FDA Received Date08/12/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/18/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-