• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON, DICKINSON & CO. (SPARKS) BD BBL¿ CHOCOLATE II AGAR (GC II AGAR WITH HEMOGLOBIN AND ISOVITALEX¿); CULTURE MEDIA, FOR ISOLATION OF PATHOGENIC NEISSERIA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON, DICKINSON & CO. (SPARKS) BD BBL¿ CHOCOLATE II AGAR (GC II AGAR WITH HEMOGLOBIN AND ISOVITALEX¿); CULTURE MEDIA, FOR ISOLATION OF PATHOGENIC NEISSERIA Back to Search Results
Model Number 221267
Medical Device Problem Code Contamination (1120)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/28/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A device evaluation and/or device history review is anticipated, but is not complete.Upon completion, a supplemental report will be filed.
 
Event or Problem Description
It has been reported that the bd bbl¿ chocolate ii agar (gc ii agar with hemoglobin and isovitalex¿) has been found contaminated.The following has been provided by the initial reporter: it was reported by the customer that they have contamination on one plate in lot 1301556.
 
Additional Manufacturer Narrative
H.6 investigation summary: complaint investigation: during manufacturing of material 221267, media is formulated and sent through a high temperature short time sterilizer to remove bioburden.The petri dishes are subjected to uv radiation to decrease bioburden.The petri dishes are filled in a positive pressure hepa filtered environment.The filled plates are cooled and immediately wrapped into sleeves to decrease the introduction of microbes.Sleeves are then packaged into cartons and then transferred to a refrigerated truck (2 to 8 degrees c) for shipment to the distributor.Bd distributors are provided with the storage guidelines for the shipping and handling of bd media of 2 to 8 degrees c in a dark place.The batch history record for batch 1301556 was satisfactory and no quality notifications were generated during manufacturing and inspection.The release testing that is performed on this product does include bioburden testing.A sample of plates are incubated at 25 degrees c and at 35 degrees c for approximately 72 hours.All bioburden testing performed on this batch was satisfactory per bd internal procedures.Affected product does not have any sterility claims; the product is tested for bioburden prior to release to ensure that it conforms to product specifications.However, this does not ensure that the end-user will not receive a contaminated plate.The complaint history was reviewed, and no other complaints have been taken on batch 1301556.Retention samples from batch 1301556 were not available for inspection.Two photos were received for investigation.One photo shows the bottom of a plate from batch 1301556 (time stamp 1715) with the plate print featured for batch verification.The other photo shows the agar surface of a plate with one surface bacterial colony.No return samples were received for investigation.This complaint can be confirmed.A trend was identified for contamination and investigation found opportunities for bioburden reduction in the manufacturing process.A capa (corrective and preventative actions) has been initiated and involves implementing additional cleaning events and evaluation of manufacturing procedures focused on in-process bioburden reduction.Additional trainings are planned with an ongoing training review for cleaning processes.Bd will continue to trend complaints for contamination.Risk management file review assessed the potential risk for the defect as severity 1 per baltrmppmgenpuraph, rev 02, id 6.13.
 
Event or Problem Description
It has been reported that the bd bbl¿ chocolate ii agar (gc ii agar with hemoglobin and isovitalex¿) has been found contaminated.The following has been provided by the initial reporter: it was reported by the customer that they have contamination on one plate in lot 1301556.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD BBL¿ CHOCOLATE II AGAR (GC II AGAR WITH HEMOGLOBIN AND ISOVITALEX¿)
Common Device Name
CULTURE MEDIA, FOR ISOLATION OF PATHOGENIC NEISSERIA
Manufacturer (Section D)
BECTON, DICKINSON & CO. (SPARKS)
7 loveton circle
sparks MD 21152
Manufacturer (Section G)
BECTON, DICKINSON & CO. (SPARKS)
7 loveton circle
sparks MD 21152
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key13260550
Report Number1119779-2022-00027
Device Sequence Number20023775
Product Code JTY
UDI-Device Identifier10382902212673
UDI-Public10382902212673
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
N/A
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date02/22/2022
Device Model Number221267
Device Catalogue Number221267
Device Lot Number1301556
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/28/2021
Supplement Date Received by Manufacturer11/01/2022
Initial Report FDA Received Date01/14/2022
Supplement Report FDA Received Date11/10/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/28/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-