Model Number FV438T
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code Hydrocephalus (3272)
Date of Event 12/22/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Manufacturing site evaluation: investigation on-going.Should relevant additional information / investigation results become available, a supplemental medwatch report will be submitted.
 
Event or Problem Description
It was reported that a progav (part # (b)(4)) was implanted during a procedure performed on (b)(6) 2019.According to the complainant, the shunt system was believed to be difficult to adjust.The patient underwent a revision procedure on (b)(6) 2021.The complaint device has not yet been returned to the manufacturer for evaluation.No patient complications were reported as a result of the revision procedure.Patient data: age: (b)(6), weight: 16 kilograms (kg), height: (b)(6) (cm), gender: unknown.
 
Additional Manufacturer Narrative
Investigation: visual inspection: the following observations were made during the visual inspection: -no abnormalities.Measurement of surface of the housing: deformation detected: yes - measured value: -0.037 mm [tolerance (0 ± 0.02mm)].Too much pressure on the valve, for example through the adjustment tool or resulting from a fall or hit to the head of the patient, can compromise the integrity of the valve.Permeability test: the test showed that the progav is permeable.Adjustability test: the progav was found to be not adjustable to specifications.Braking force and brake function test: the braking force test indicates that the brake function of the progav is present.Due to the non-adjustability of the valve, an investigation of the braking force was not possible.Internal inspection of product: in order to verify whether the investigated valve was compromised by the known risks of hydrocephalus therapy, e.G.By a build-up of natural substances (protein, blood, or tissue particles) in the cerebrospinal fluid, we have dismantled the valve.After dismantling of the valve, some deposits were found in progav.Results: based on our investigation results, we can identify a "non-adjustability" of the valve.We are assuming that the deposits detected within the valve have caused the functional impairment.Deposits caused by natural substances in the cerebrospinal fluid, such as protein, blood or tissue particles, are among the known and unavoidable risks and side effects of hydrocephalus therapy.Small amounts of non-visible deposits/proteins might compromise the integrity of the valve.We can exclude a defect at the time of release.The valve met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.From our point of view, no further regulatory actions are required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGAV SYSTEM PED.WITH PRECHAMBER
Common Device Name
HYDROCEPHALUS MANAGEMENT
Manufacturer (Section D)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM  14469
Manufacturer (Section G)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM   14469
Manufacturer Contact
joerg knebel
ulanenweg 2
potsdam, 14469
GM   14469
MDR Report Key13260881
Report Number3004721439-2021-00368
Device Sequence Number1701208
Product Code JXG
UDI-Device Identifier04041906132514
UDI-Public4041906132514
Combination Product (Y/N)N
Initial Reporter CountryRS
PMA/510(K) Number
K103003
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/14/2024
Device Model NumberFV438T
Device Catalogue NumberFV438T
Device Lot Number20041642
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/20/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/24/2021
Supplement Date Received by Manufacturer01/25/2022
Initial Report FDA Received Date01/14/2022
Supplement Report FDA Received Date02/18/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/23/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age4 YR
Patient SexUnknown
Patient Weight16 KG
-
-