| Model Number |
120-500 |
| Medical Device Problem Code |
Device Appears to Trigger Rejection (1524)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.The event of capsular contracture is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: baker grade unknown.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported left side "exchange of textured device due to patient's concern with product".Healthcare professional additionally reported "capsular ctx", baker grade not specified.Device has been explanted and replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: visual analysis of the returned device identified wear abrasion, deformation, yellow particles material was found on the shell and bubbles.A weight test of the device was verified and the device was within specification.Cloudy after autoclave disinfection process in the gel.Based on the device analysis the final assessment is: no issues found related with the manufacturing.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported left side "exchange of textured device due to patient's concern with product".Healthcare professional additionally reported "capsular ctx", baker grade not specified.Device has been explanted and replaced.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported left side "exchange of textured device due to patient's concern with product".Healthcare professional additionally reported "capsular ctx", baker grade not specified.Device has been explanted and replaced.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported left side "exchange of textured device due to patient's concern with product".Healthcare professional additionally reported "capsular ctx", baker grade not specified.Device has been explanted and replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional, changed, and/or corrected data: d6a, g1.
|
| |
|