EXACTECH, INC. LOGIC CR TIB INSERT SLOPE++, SZ 2, 13MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED
|
Back to Search Results |
|
| Model Number |
LOGIC CR TIB INSERT SLOPE++, SZ 2, 13MM |
| Medical Device Problem Codes |
Naturally Worn (2988); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Inflammation (1932); Pain (1994); Scar Tissue (2060); Synovitis (2094); Osteolysis (2377); Ambulation or Postural Difficulties (2544); Balance Problems (4401); Swelling/ Edema (4577)
|
| Date of Event |
11/01/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Additional information, including the product investigation, will be submitted within 30 days of receipt.
|
| |
|
Event or Problem Description
|
|
As reported, a (b)(6) woman who had a left total knee replacement done for medial compartment oa back in 2018.Patient did exceedingly well and achieved a range of motion (rom) of 0-138 degrees.Patient resumed full activities including going to the gym and doing low impact exercises and returned to her job as a dental assistant.Patient visit surgeon last month complaining of worsening proximal tibial pain and swelling.No history of trauma.No constitutional symptoms to suggest infection.Patient exam showed that she still maintained her rom but clearly had a 1-2+ effusion without erythema nor unusual warmth.Her x-rays from (b)(6) 2021 were compared to her x-rays from (b)(6) 2018 and this clearly showed radiolucencies around the medial femoral condyle, the medial tibial plateau under the baseplate and around the keel of the tibial baseplate as well.Furthermore, of greater concern was the fact that her poly thickness had decreased by approximately 50%.Patient has subsequently had a 3 phase bone scan which shows increased activity in all 3 of those same areas consistent with loosening and a mavric mri which shows a stress injury under her medial tibial plateau.Her serology has been negative for infection including a normal crp and sed rate and wbc.She has a joint aspiration pending but again clinically has no sign of infection.Patient is on my list of patients who have the greater than 5 year shelf life of optetrak poly and surgeon think patient needs a revision of her tibial component and possibly her femoral component as well.
|
| |
|
Additional Manufacturer Narrative
|
|
After further review of additional information received the following sections d1, d4, d10, g3, g4, g6, h2 and h4 have been updated accordingly.Section d1: brand name section d4: model number, catalog number, serial number, expiration date, unique identifier (udi) # section d10: concomitant medical products logic cr femoral cem, left sz 2 (cat# 02-010-03-0220 / serial# (b)(6)) lgc tibial fit tray cem sz 2f / 3t (cat# 02-012-45-2030 / serial# (b)(6)) advanced patella 29m 3 peg implant (cat# 200-07-29 / serial# (b)(6)) 3" drill bit, mod.Hex 2 pack (cat# 201-78-89 / serial# (b)(6)) section g4: pma/510(k)number section h4: device manufacture date.Additional information, including the product investigation, will be submitted within 30 days of receipt.
|
| |
|
Additional Manufacturer Narrative
|
|
After further review of additional information.(h3) the revision reported was likely the result of both patient-related conditions and an insufficient bond between the femoral component and the bone, which led to aseptic (non-infected) femoral loosening.The loosening may have led to instability and an increase in cement and bone particles in the joint space, which resulted in prosthesis wear and osteolysis.Additionally, a contributing factor to the wear and delamination may have been the result of being packaged in a non-conforming vacuum bag for more than five years.However, this cannot be confirmed as the device were not available for evaluation.The most probable root cause associated with the reported event of "prosthesis wear¿ is associated with material damage to a surface, usually involving progressive loss or displacement of material, due to relative motion between that surface and a contacting substance or substances.Also, the most probable root cause associated with the reported event of "loosening¿ is associated with weakened integration of the device at the bone-implant interface due to loss of fibrous and/or bony tissue and leading to compromised anchorage of the device.Furthermore, the most probable root cause associated with the reported event of "osteolysis¿ is associated with destruction or disappearance of bone tissue likely related to weakened integration of an implant at the bone-implant interface.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation has been re-opened based on receipt of additional information.Fields d6a/d6b, and g3 have been updated accordingly.Additional information, including the investigation results, will be submitted within 30 days of receipt.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information- h7 & h9.There is no requirement for additional investigation.These devices are used for treatment not diagnosis.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional and/or corrected information.The following sections were updated/corrected: h6 mdr section codes updated/corrected: a, b, c, d, e, f, g the revision reported was likely the result of both patient-related conditions and an insufficient bond between the femoral component and the bone, which led to aseptic (non-infected) femoral loosening.The loosening may have led to instability and an increase in cement and bone particles in the joint space, which resulted in prosthesis wear and osteolysis.Additionally, a contributing factor to the wear and delamination may have been the result of being packaged in a non-conforming vacuum bag for more than five years.However, this cannot be confirmed as the device were not available for evaluation if any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|