CHRISTOPH MIETHKE GMBH & CO KG M.BLUE 15 VALVE; HYDROCEPHALUS MANAGEMENT Back to Search Results
Model Number FX803T
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code Hydrocephalus (3272)
Date of Event 10/21/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a m.Blue 15 valve (part # fx803t) was implanted during a procedure performed on (b)(6) 2019.According to the complainant, the shunt system was believed to be difficult to adjust.The patient underwent a revision procedure on (b)(6) 2021.The complaint device was returned to the manufacturer for evaluation.No patient complications were reported as a result of the revision procedure.Age: (b)(6), height: 163 centimeters (cm), weight: (b)(6), gender: male.
 
Additional Manufacturer Narrative
Manufacturing and quality control data.The m.Blue® was manufactured by a qualified employee in june 2019.Deviations during assembly did not occur.The valve was sterilized by miethke and released for shipment after final inspection.The m.Blue® has a fixed pressure of 15 cmh2o in the horizontal position and a normal pressure range of 0 to 40 cmh2o in the vertical position.The valve was inspected as article (b)(4).All parameters have been inspected and approved during the manufacturing process.All parameters have been assessed as meeting specifications.Visual inspection: no significant deformations or damage of the valve were detected during the visual inspection.Permeability test: a permeability test has shown that the valve is permeable.Adjustment test: the m.Blue® valve was tested and is adjustable to all specified pressures.Braking force and brake function test: the brake functionality test has shown that the brake function is fully operational and the braking force is within the given tolerances.Internal inspection : after dismantling of the valve, minimal deposits were found in m.Blue®.Results: based on our investigation, we are unable to substantiate the clinical claim of non adjustability.At the time of our investigation the valve was inside the tolerances.However, it is possible that the deposits observed inside the valve could have caused the malfunction in the past.As described in scientific literature, the problem encountered is one of the known, inevitable risks of hc-therapy by shunt implants.We can exclude a defect at the time of release.The valve met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.No further regulatory actions are required from our point of view.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
M.BLUE 15 VALVE
Common Device Name
HYDROCEPHALUS MANAGEMENT
Manufacturer (Section D)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM  14469
Manufacturer (Section G)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM   14469
Manufacturer Contact
joerg knebel
ulanenweg 2
potsdam, 14469
GM   14469
MDR Report Key13261499
Report Number3004721439-2021-00362
Device Sequence Number4760944
Product Code JXG
UDI-Device Identifier04041906504427
UDI-Public4041906504427
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K192266
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 01/14/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/05/2024
Device Model NumberFX803T
Device Catalogue NumberFX803T
Device Lot Number20042328
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/30/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/30/2021
Initial Report FDA Received Date01/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/07/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age29 YR
Patient SexMale
Patient Weight115 KG
-
-