• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AMO PUERTO RICO MFG. INC. TECNIS IOL; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMO PUERTO RICO MFG. INC. TECNIS IOL; INTRAOCULAR LENS Back to Search Results
Model Number PCB00
Medical Device Problem Codes Difficult to Fold, Unfold or Collapse (1254); Inaccurate Delivery (2339)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/16/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Initial reporter telephone number: (b)(6).The lens was not returned for evaluation as it remains implanted.Therefore, a failure analysis of the complaint device could not be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review and possible product return and evaluation, if there is any further relevant information a supplemental medwatch will be filed.Per eu regulation (general data protection regulation), patient identifiers were not collected or recorded and therefore are not available.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
During implantation, the last push was given and the trailing haptic was not folded onto the optic and instead unfolded.The doctor implanted the lens normally after using a mac pherson pincer to get the trailing haptic out of the cartridge.The inserter is available while the lens remains implanted.No delay and no medical or surgical interventions were reported and no further details provided.This emdr report is for the second reported event.Separate reports will be submitted for the other two events.
 
Additional Manufacturer Narrative
Additional information: section d9: device available for evaluation? yes.Section d9: returned to manufacturer? yes.Section d9: date returned to manufacturer: 18th february 2022.Section h3: device evaluated by manufacturer? yes.Device evaluation: visual inspection under magnification revealed that the handpiece was received with the plunger rod locked and in the advanced position.No issues were identified with the external assembly however, viscoelastic residue was observed on the outside of the cartridge.The handpiece was disassembled and the assembly was inspected, presenting with no issues.No complaint lens was not received as part of this return as it is still implanted.Therefore, no product evaluation of the lens could be performed.The complaint issue could not be confirmed, and no product deficiency could be identified.The customer experience may be attributed to an excessive amount of ophthalmic viscoelastic device (ovd), suggested by the presence of viscoelastic residue on the outside of the cartridge as well as the trace amounts found inside of the lower body assembly.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
AMO PUERTO RICO MFG. INC.
road 402 north, anasco ind. pk
anasco PR 00610
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key13262429
Report Number3012236936-2022-00123
Device Sequence Number1546759
Product Code HQL
UDI-Device Identifier05050474558342
UDI-Public(01)05050474558342(17)240715
Combination Product (Y/N)N
Initial Reporter CountryNL
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/15/2024
Device Model NumberPCB00
Device Catalogue NumberPCB0000240
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/23/2021
Supplement Date Received by Manufacturer03/24/2022
Initial Report FDA Received Date01/15/2022
Supplement Report FDA Received Date04/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/15/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-