| Model Number |
MR8-AD03 |
| Medical Device Problem Code |
Device Remains Activated (1525)
|
| Health Effect - Clinical Codes |
Hemorrhage/Blood Loss/Bleeding (1888); Perforation (2001)
|
| Date of Event |
12/17/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that during deep brain stimulator placement, the midas rex mr8 pneumatic drill with a mr8-ad03 perforator driver and disposable codman perforator failed to stop spinning and drilled into the patient's brain.On follow-up, it was reported that there was no staff impact and the patient was injured with excessive bleeding, which would be the case in normal operation.The case was delayed for about an hour as the surgeons got the bleeding under control.The patient has been discharged and was okay.
|
| |
|
Additional Manufacturer Narrative
|
|
No conclusion can be drawn.No evaluation was performed, as the device was not returned.If the device is returned in the future, product analysis may be performed.The device user manual warnings section includes instructions to check the device for damage before use.If damage is found, the device should not be used.We will continue to track and trend this complaint type.Other relevant device(s) are: product id: pm800, serial/lot #: (b)(4), udi#: (b)(4).If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Report not confirmed.Evaluation could not reproduce the reported malfunction of continues to run.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|