| Model Number |
654301 |
| Medical Device Problem Code |
Contamination (1120)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/29/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device has been received for evaluation, however, investigation is not yet complete.
|
| |
|
Event or Problem Description
|
|
It was reported that an eyelash was identified in the bladder irrigation tubing of a ltxfr nv tur y-set.The package was sealed prior to the customer finding the eyelash.There was no patient involvement, no adverse event and no harm.
|
| |
|
Additional Manufacturer Narrative
|
|
One new list number 065430403, t-u-r y-set, nonvented, 96 inch; lot number 5455555.The complaint of hair in the tubing/fluid path could not be confirmed on the one returned unit, however, hair was observed inside the packaging of the returned unit.No other anomalies or damages were observed.The probable cause is due to an error during gowning at the manufacturing site.Lot history review was performed and no nonconformities were identified that may have contributed to the reported complaint.
|
| |
|