| Model Number |
5820013130 |
| Medical Device Problem Codes |
Fracture (1260); Dull, Blunt (2407)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/20/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a spinal surgical procedure, the tip of the bur broke.It was also reported that there were no adverse consequences and no delays as a result of this event.It was further reported that the procedure was completed successfully.
|
| |
|
Additional Manufacturer Narrative
|
|
A follow up report will be filed once the quality investigation is complete.Awaiting device return.
|
| |
|
Additional Manufacturer Narrative
|
|
B5: it was initially reported that the bur broke, upon evaluation, the diamond grit on the bur head was worn, this is not a reportable event.H6: the quality investigation is complete.
|
| |
|
Event or Problem Description
|
|
It was reported that during a spinal surgical procedure, the tip of the bur broke.It was also reported that there were no adverse consequences and no delays as a result of this event.It was further reported that the procedure was completed successfully.It was further reported that following evaluation of the returned bur, it not break the diamond grits were observed to be completely worn off.
|
| |
|