MAUDE Adverse Event Report: STRYKER INSTRUMENTS-A DIVISION OF STRYKER CORP 3.0MM X-COARSE DIAMOND; DRILLS, BURRS, TREPHINES & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER INSTRUMENTS-A DIVISION OF STRYKER CORP 3.0MM X-COARSE DIAMOND; DRILLS, BURRS, TREPHINES & ACCESSORIES Back to Search Results
Model Number 5820013130
Medical Device Problem Codes Fracture (1260); Dull, Blunt (2407)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/20/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a spinal surgical procedure, the tip of the bur broke.It was also reported that there were no adverse consequences and no delays as a result of this event.It was further reported that the procedure was completed successfully.
 
Additional Manufacturer Narrative
A follow up report will be filed once the quality investigation is complete.Awaiting device return.
 
Additional Manufacturer Narrative
B5: it was initially reported that the bur broke, upon evaluation, the diamond grit on the bur head was worn, this is not a reportable event.H6: the quality investigation is complete.
 
Event or Problem Description
It was reported that during a spinal surgical procedure, the tip of the bur broke.It was also reported that there were no adverse consequences and no delays as a result of this event.It was further reported that the procedure was completed successfully.It was further reported that following evaluation of the returned bur, it not break the diamond grits were observed to be completely worn off.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3.0MM X-COARSE DIAMOND
Common Device Name
DRILLS, BURRS, TREPHINES & ACCESSORIES
Manufacturer (Section D)
STRYKER INSTRUMENTS-A DIVISION OF STRYKER CORP
1941 stryker way
portage MI 49002
Manufacturer (Section G)
STRYKER INSTRUMENTS-IRELAND
instruments division
carrigtwohill bus. & tech park
carrigtwohill NA
EI   NA
Manufacturer Contact
una barry
instruments division
carrigtwohill bus. & tech park
carrigtwohill NA
EI   NA
214532900
MDR Report Key13266027
Report Number3015967359-2022-00063
Device Sequence Number16569366
Product Code HBE
UDI-Device Identifier04546540374288
UDI-Public04546540374288
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K143320
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/01/2026
Device Model Number5820013130
Device Catalogue Number5820013130
Device Lot Number21175017
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/21/2021
Supplement Date Received by Manufacturer03/23/2022
Initial Report FDA Received Date01/17/2022
Supplement Report FDA Received Date03/25/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/24/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-