• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER TRAUMA KIEL UNKNOWN 10*320 T2 ALPHA FEMORAL NAIL PF LEFT; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER TRAUMA KIEL UNKNOWN 10*320 T2 ALPHA FEMORAL NAIL PF LEFT; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Catalog Number UNK_KIE
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Non-union Bone Fracture (2369)
Event Date 12/13/2021
Event Type  Injury  
Manufacturer Narrative
This complaint has been generated based on findings discovered during post market surveillance pmcf study review.The alleged nonunion for which revision surgery is scheduled included in the study could not be confirmed, since the device was not returned for evaluation and no other additional information will be made available.More detailed information about the patient medical history, the event circumstances, radiographs and the involved device(s) must be available in order to determine the root cause.If any additional information becomes available, the investigation will be reopened and re-evaluated accordingly.Device disposition is unknown.
 
Event Description
The manufacturer became aware of a pmcf final study report that was conducted by university of louisville in the usa.The title of this report is ¿a retrospective data collection of the treatment of femoral fractures with the femoral nail greater trochanter (pf) of the t2 alpha femur antegrade gt/pf nailing, which is associated with the stryker ¿t2 alpha femur antegrade gt/pf nailing¿ system.This report includes analysis of the clinical data that was collected on 28 patients, the cases in this study range from march 2019 through august 2020.During the review of the literature, it was not possible to establish a specific device details, patient information, and at this time no additional device information is available.It was reported that one patient experienced nonunion for which revision surgery is scheduled.The report states: ¿patient j (1 device related event): the 4th nonunion case was (b)(6) year old female, bmi (b)(6), nonsmoker, with history of alcohol use.Her injury was due to mva and was classified as closed 32-c3 fracture.The devise used for the initial fixation surgery was 10*320 t2 alpha femoral nail pf left, with recon locking mode.The nonunion was diagnosed 380 days post initial surgery and was advised a revision surgery and will be scheduled for repair of nonunion.¿.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM  D-24232
Manufacturer (Section G)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM   D-24232
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key13266265
MDR Text Key284901095
Report Number0009610622-2022-00020
Device Sequence Number1
Product Code HSB
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Study
Reporter Occupation Physician
Type of Report Initial
Report Date 01/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/17/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Catalogue NumberUNK_KIE
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/19/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age49 YR
Patient SexFemale
-
-