As reported to customer relations via phone conversation "after deployment (and dilation with a balloon) the device migrated from the iliac into the ivc." the physician had to make an additional access site to use a balloon to the pull stent more into the ivc.Placed another balloon (competitive company) distal to the stent to overlap-to hold it in place.Patient outcome: did any unintended section of the device remain inside the patient¿s body? no.Please describe the object and how it was retrieved: did the patient experience a delay or require any additional procedure due to this occurrence? no.Please specify delay or any additional procedure(s) and provide details: has the complainant reported any adverse effects on the patient due to this occurrence? no.Has the complainant reported that the product caused or contributed to the adverse effects? no.Please specify adverse effects and provide details.Patient/event info - notes: are images of the device or procedure available? n/a, yes, no? yes.Did the patient have pre-existing conditions? n/a, yes, no? yes.If yes, please specify: asku.Please describe the native state of the vessel (i.E.Was the anatomy tortuous? asku was the vessel fibrotic?) n/a, tortuous, calcified, fibrotic, other.If other, please specify.Was a stent previously placed during previous procedures? n/a, yes, no? was the device used percutaneously? n/a, yes, no? where on the patient was the percutaneous access site? was the access site jugular or femoral? n/a, jugular, femoral other.If other, please specify.What disease pathology was being treated? may thurner, acute or chronic obstruction, restenosis, other? if other, please specify was the lesion approached via contralateral or ipsilateral? n/a, contralateral, ipsilateral.Was pre-dilation performed ahead of placement of the stent? n/a, yes, no? what was the target location for the stent? details of access sheath used (name, fr size, length)? was the device flushed through both flushing ports before the procedure, as per ifu? n/a, yes, no? details of the wire guide used (name, diameter, hydrophylic)? was resistance encountered when advancing the wire guide to the target location? n/a, yes, no? was resistance encountered when advancing the delivery system to the target location? n/a, yes, no? if resistance was met, how did the physician address this? did the tip of the delivery system cross the target location? n/a, yes, no? did the user pull the handle towards the hub during deployment, per ifu? n/a, yes, no? did the user push the hub during deployment? n/a, yes, no? did the user remove slack in the delivery system before deployment, per ifu? n/a, yes, no? was the stent deployed smoothly /without resistance? n/a, yes, no? was the stent fully deployed in the patient? n/a, yes, no? was the stent fully deployed before removing the delivery system from the patient? n/a, yes, no? was post dilation performed after the placement of the stent? n/a, yes, no? was the delivery system damaged/kinked/twisted during deployment? n/a, yes, no? what intervention (if any) was required? was the secondary intervention performed during the same procedure as the device failure or was it scheduled for another day? n/a, same procedure, another day.Were any other defects (other than the complaint issue) observed on the delivery system prior to return (e.G.Kink)? n/a, yes, no? please specify if yes.Are images of the device or procedure available? yes.Did the patient have pre-existing conditions? no.Please describe the native state of the vessel (i.E.Was the anatomy tortuous? was the vessel fibrotic?) tortuous, compressed.Was a stent previously placed during previous procedures? yes, left stent.Was the device used percutaneously? yes.Where on the patient was the percutaneous access site? neck right internal jugular.Was the access site jugular or femoral? internal jugular.What disease pathology was being treated? right external compression 57.4%.Was the lesion approached via contralateral or ipsilateral? neck right internal jugular.Was pre-dilation performed ahead of placement of the stent? no.What was the target location for the stent? right eiv.Details of access sheath used (name, fr size, length) 10fr boston scientific 23cm.Was the device flushed through both flushing ports before the procedure, as per ifu? yes.Details of the wire guide used (name, diameter, hydrophylic)? zipwire hydrophilic wire 260cm.035.Was resistance encountered when advancing the wire guide to the target location? no.Was resistance encountered when advancing the delivery system to the target location? no.If resistance was met, how did the physician address this? no resistance.Did the tip of the delivery system cross the target location? yes.Did the user pull the handle towards the hub during deployment, per ifu? yes.Did the user push the hub during deployment? no.Did the user remove slack in the delivery system before deployment, per ifu? yes.Was the stent deployed smoothly /without resistance? yes.Was the stent fully deployed in the patient? yes.Was the stent fully deployed before removing the delivery system from the patient? yes.Was post dilation performed after the placement of the stent? yes.Was the delivery system damaged/kinked/twisted during deployment? no.What intervention (if any) was required? stent retraction.Was the secondary intervention performed during the same procedure as the device failure or was it scheduled for another day? same procedure.Were any other defects (other than the complaint issue) observed on the delivery system prior to return (e.G.Kink)? no.
|