• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR XIENCE SKYPOINT DRUG ELUTING CORONARY STENT DELIVERY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR XIENCE SKYPOINT DRUG ELUTING CORONARY STENT DELIVERY SYSTEM Back to Search Results
Model Number 1804400-12
Medical Device Problem Codes Difficult to Remove (1528); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/27/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device is not returning for analysis.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event or Problem Description
It was reported that the procedure was to treat a de novo, 90% stenosed, concentric, mildly calcified, tortuous right coronary artery.During preparation of the 4.0x12 mm xience skypoint stent delivery system (sds) there was difficulty removing the protective sheath.Additionally the stent was noted to have moved on the balloon.There was no device use or patient involvement.A 4.0x15 mm xience skypoint sds was used to complete the procedure.There was no reported clinically significant delay in the procedure.No additional information was provided.
 
Additional Manufacturer Narrative
The device was not returned for evaluation.A review of the lot history record revealed no manufacturing nonconformities that would have contributed to this complaint.Additionally, a review of the complaint history did not indicate a lot specific quality issue.The investigation determined the reported difficulties appear to be related to the operational context of the procedure.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Event or Problem Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE SKYPOINT DRUG ELUTING CORONARY STENT DELIVERY SYSTEM
Common Device Name
DRUG ELUTING CORONARY STENT DELIVERY SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005718570 (P099)
cashel road
clonmel tipperary
EI  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key13266810
Report Number2024168-2022-00556
Device Sequence Number4726313
Product Code NIQ
UDI-Device Identifier08717648233463
UDI-Public08717648233463
Combination Product (Y/N)Y
Initial Reporter CountryJA
PMA/510(K) Number
P110019
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/09/2023
Device Model Number1804400-12
Device Catalogue Number1804400-12
Device Lot Number0090341
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 12/27/2021
Supplement Date Received by Manufacturer01/19/2022
Initial Report FDA Received Date01/17/2022
Supplement Report FDA Received Date02/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/10/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-