| Model Number |
1012273-12 |
| Medical Device Problem Codes |
Difficult to Remove (1528); Material Rupture (1546); Difficult to Advance (2920)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/27/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the procedure was to treat the heavily calcified, moderately tortuous, 90% stenosed and concentric lesion in the distal right coronary artery (rca).The 2.75x12 mm trek balloon dilatation catheter (bdc) was advanced with resistance with the anatomy and was inflated once to 6 atmospheres and ruptured.There was resistance with the anatomy noted during independent removal.A new trek bdc was used to complete the procedure.There were no adverse patient effects and there was reported clinically significant delay in the procedure.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.The investigation determined the reported complaints appear to be related to operational context.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.
|
| |
|