An interventional radiology technician reported an issue with a mini stick max 4f x 10cm stiff.018 ni/tu echo 2.75." during a procedure, when the dr.Went to remove the inner dilator from the sheath the dilator broke off at the hub.The remaining dilator was in the sheath so the dr.Used a snare to remove it from the sheath.No part of the dilator broke off in the patient and the procedure continued.There was no report of the patient experiencing any adverse effects, harm, or require medical intervention as a result of this incident.
|
|
As the reported device was not returned, angiodynamics is unable to perform a device evaluation.The supplier of this component was made aware of this event via (b)(4).The customer's reported complaint description cannot be confirmed since no dilator sample was returned for evaluation.Without receiving product for evaluation, we are unable to definitively determine a root cause for this incident.A review of the device history records was performed for the indicated lots for any deviations related to the reported failure mode of the complaint.The review confirms that the lots met all material, assembly and performance specifications; i.E.No ncr associated with reported failure mode.Labeling review: direction for use {dfu} is provided with this mini-stick device.The failure mode of device fracture and embolization is cautioned as potential adverse event.No sample was returned for evaluation so it cannot be determined if device was used in a manner inconsistent with labeling.Adverse events: guidewire shearing, fracture or embolization.Operational instructions: 1.Prior to insertion, flush all components with saline or heparinized/saline.2.Gain percutaneous access with the 21 g (0.9 mm) vascular introducer needle.3.Advance the 0.018 inch (0.46 mm) guidewire through the needle.Caution: the guidewire should not be withdrawn through the introducer needle.If the guidewire must be withdrawn while still inside the needle, simultaneously remove both the needle and the wire as a unit to prevent the needle from damaging the guidewire.4.Withdraw the introducer needle, leaving the 0.018 inch (0.46 mm) guidewire in place.5.Advance the coaxial assembly over the 0.018 inch (0.46 mm) guidewire.6.Simultaneously remove the dilator and 0.018 inch (0.46 mm) guidewire, while holding the sheath portion of the assembly in place.7.Advance a 0.035 inch (0.89 mm) or 0.038 inch (0.97 mm) guidewire into the sheath.8.Remove the sheath, leaving the 0.035 inch (0.89 mm) or 0.038 inch (0.97 mm) wire.A review of the angiodynamics complaint system noted no adverse trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.Reference (b)(4).
|