| Model Number |
359529 |
| Medical Device Problem Code |
Extracardiac stimulation (1463)
|
| Health Effect - Clinical Code |
Discomfort (2330)
|
| Date of Event |
01/11/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Patient reports discomfort at implant site (abdominal) when he can tell device is pacing.Testing in office suggests pocket stimulation due to unipolar pacing and high output, as well as patient discomfort with pacing.Output was lowered and rate response turned off to reduce unnecessary pacing per physician.Should additional information be received, this file will be updated.
|
| |
|
Additional Manufacturer Narrative
|
|
The device is currently not available for analysis.No conclusion can be drawn at this time.No additional information is available at the moment.The file is closed.The investigation will be re-opened should additional data become available.
|
| |
|