| Model Number |
ETLW1628C124E |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
01/15/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
An endurant limb was implanted during an endovascular procedure for the treatment of a 90mm abdominal aortic aneurysm. it was reported approximately 7 years post the index procedure, the patient presented emergently.A type ib was identified in the left limb (etlw1628c124e).Intervention was completed where the physician extended the left limb to the left external iliac using a 16/16/156 limb from the flow divider, then extending into the left external iliac with a 16/16/93 limb and ballooned with a reliant.The endoleak was resolved and patient was doing well after the procedure. as per the physician the cause of the event was due to the patients anatomy as a result of continued disease progression of the common iliac arteries.No additional clinical sequelae was provided and the patient is fine.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|