• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW ORTHOPAEDICS AG PLUS SL MIA DBL OFFSET ADAPT. RT 60/25MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER, CEMENTED OR NON-POROUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW ORTHOPAEDICS AG PLUS SL MIA DBL OFFSET ADAPT. RT 60/25MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER, CEMENTED OR NON-POROUS Back to Search Results
Model Number 75004613
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/04/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event or Problem Description
It was reported that, during a thr surgery, the end of a plus sl mia dbl offset adapt.Rt 60/25mm, that connects to the strike pad, broke off.It is unknown if any pieces fell inside of the patient.Surgery was resumed, without any delay, with a back-up device.Patient was not injured as consequence of this problem.
 
Additional Manufacturer Narrative
H3, h6: it was reported that, during a total hip replacement surgery, the end of a plus sl mia double offset adapter right 60/25mm, that connects to the strike pad, broke off.The device, used in treatment, was returned for investigation.Upon visual , the reported failure mode could be confirmed.The clutch is broken.Apart from that, the device shows normal signs of usage such as gouges and dents.A complaint history review was performed.Three additional complaints for the batch in scope were reported so far but with another reported failure mode.A review of the batch record was performed.No deviations from the standard manufacturing process which could have contributed to the reported issue were reported.The failure mode and the severity are covered in the corresponding risk management document.A review of past corrective actions was performed.No further escalation is required.There is no indication that the reported device failed to meet manufacturing specifications upon release for distribution.However, based on the performed investigations, the relationship between the reported event and the device was confirmed.This instrument is designed to hold and guide the detachable rasps for broaching.It is therefore subjected to repeated impact forces via the modular knock plate or the pneumatic woodpecker.Previous investigations demonstrated, that under specific circumstances, this device may fracture during impaction.An optimized design of the device has been released in order to reduce the occurrence of this issue.According to document "processing (cleaning, disinfection and sterilization) of instruments from smith & nephew orthopaedics ag" (lit.N°03389-en 1363 v3 11/19), all instruments must be inspected and controlled for proper functioning after cleaning/disinfection.This version of the device will be monitored for similar issues.The returned device will be discarded.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PLUS SL MIA DBL OFFSET ADAPT. RT 60/25MM
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER, CEMENTED OR NON-POROUS
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS AG
schachenallee 29
aarau CH-50 00
SZ  CH-5000
Manufacturer (Section G)
SMITH & NEPHEW ORTHOPAEDICS AG
schachenallee 29
aarau CH-50 00
SZ   CH-5000
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key13280623
Report Number9613369-2022-00007
Device Sequence Number17381239
Product Code LZO
UDI-Device Identifier07611996085471
UDI-Public07611996085471
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K123598
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number75004613
Device Catalogue Number75004613
Device Lot Number1502
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/11/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/04/2022
Supplement Date Received by Manufacturer02/03/2022
Initial Report FDA Received Date01/18/2022
Supplement Report FDA Received Date02/04/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/28/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-