• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION GLADIATOR ELITE; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION GLADIATOR ELITE; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL Back to Search Results
Model Number 24686
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/23/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that balloon rupture occurred.The stenosed target lesion was located in an arteriovenous fistula (avf).A 7.0 x80, 75cm gladiator elite balloon catheter was advanced for dilatation.However, during the procedure, the balloon burst.The procedure was completed with another of same device.There were no patient complications reported and the patient status was stable.
 
Event or Problem Description
It was reported that balloon rupture occurred.The stenosed target lesion was located in an arteriovenous fistula (avf).A 7.0 x80, 75cm gladiator elite balloon catheter was advanced for dilatation.However, during the procedure, the balloon burst.The procedure was completed with another of same device.There were no patient complications reported and the patient status was stable.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: a gladiator device, 20 atm was returned for analysis, the following attributes were examined.A visual examination identified that the balloon was not folded which indicates that the balloon was subjected to positive pressure.Blood was present in the balloon which is indicative of a leak.The rated burst pressure for this device is 20 atmospheres as per gladiator elite specification.The returned device was attached to an encore inflation unit and positive pressure was applied and liquid was observed to be leaking from a balloon pinhole located approximately 30mm distal of the proximal marker band.A visual and microscopic examination observed no damage to the tip of the device.A visual and tactile examination identified multiple shaft kinks.A visual examination found no issues or damage to the markerbands of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GLADIATOR ELITE
Common Device Name
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key13281050
Report Number2134265-2022-00339
Device Sequence Number4944638
Product Code LIT
UDI-Device Identifier08714729809838
UDI-Public08714729809838
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K132810
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/23/2023
Device Model Number24686
Device Catalogue Number24686
Device Lot Number0026407407
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/18/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 01/10/2022
Supplement Date Received by Manufacturer01/27/2022
Initial Report FDA Received Date01/18/2022
Supplement Report FDA Received Date02/02/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/23/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age65 YR
Patient SexFemale
Patient Weight58 KG
-
-