Model Number 71358218
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Joint Dislocation (2374)
Date of Event 12/26/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that, after a thr performed on (b)(6) 2021.The patient experienced recurrent anterior dislocation of the right hips; therefore, a revision surgery was performed on (b)(6) 2021 to revise the r3 multi hole acetabular shell 54mm, or3o dual mobility liner 42/54 and the or3o dm xlpe insert 28/42.Surgical outcome was successful.No further information is available.
 
Additional Manufacturer Narrative
Internal complaint reference: case-(b)(4).
 
Additional Manufacturer Narrative
H10: the device was not returned for evaluation and the reported event could not be confirmed.The clinical/medical investigation concluded that, no relevant supporting clinical information has been provided to assist with a clinical investigation.Per e-mail communication, the subject stated, ¿that the dislocation first occurred a few days after surgery but resolved on its own.¿ it was also noted that the patient is recovering.Based on insufficient information, a thorough clinical assessment cannot be performed at this time.Should any additional clinical information be provided this complaint will be re-evaluated.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history for the part number over the past 12 months and for the batch number based on historical data of the device did not reveal similar events for the listed device.A review of the risk management file and instructions for use documents revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Possible causes could include but not limited to traumatic injury, patient anatomy or abnormal loading of limb.The contribution of the device to the reported event could not be corroborated.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.Internal complaint reference number: case-2022-00087603-3.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OR3O DM XLPE INSERT 28/42
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key13281681
Report Number1020279-2022-00238
Device Sequence Number17368631
Product Code LPH
UDI-Device Identifier00885556712375
UDI-Public00885556712375
Combination Product (Y/N)N
Initial Reporter StateNV
Initial Reporter CountryUS
PMA/510(K) Number
K191002
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2021
Reporter Type Manufacturer
Report Source Study,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71358218
Device Catalogue Number71358218
Device Lot NumberA2021523
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/04/2022
Supplement Date Received by Manufacturer04/05/2022
Initial Report FDA Received Date01/18/2022
Supplement Report FDA Received Date04/06/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured11/26/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexFemale
-
-