MAUDE Adverse Event Report: STRYKER GMBH CANNULATED SCREW, PARTIALLY THREADED ASNIS III Ø6.5X40MM TL20MM; SCREW, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER GMBH CANNULATED SCREW, PARTIALLY THREADED ASNIS III Ø6.5X40MM TL20MM; SCREW, FIXATION, BONE Back to Search Results
Model Number 602640S
Medical Device Problem Code Difficult to Remove (1528)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/21/2021
Type of Reportable Event Malfunction
Event or Problem Description
As reported: "when the screw was removed, we tried to manually remove it by counter-rotation with a screwdriver, but the screw was buried.The back of the lower leg was expanded, and the screw was finally removed with a screwdriver while countering the screw tip with forceps." the procedure was delayed by 20 min.
 
Additional Manufacturer Narrative
The device will not be returned.If additional information becomes available, it will be provided in a supplemental report.Device discarded.
 
Additional Manufacturer Narrative
Correction: please refer to d4 lot number.The reported event could not be confirmed since the device was not returned for evaluation and no other evidences were provided.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.A review of the device history was not possible because the lot number was not communicated.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If the device is returned or if any additional information is provided, the investigation will be reassessed.
 
Event or Problem Description
As reported: "when the screw was removed, we tried to manually remove it by counter-rotation with a screwdriver, but the screw was buried.The back of the lower leg was expanded, and the screw was finally removed with a screwdriver while countering the screw tip with forceps." the procedure was delayed by 20 min.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CANNULATED SCREW, PARTIALLY THREADED ASNIS III Ø6.5X40MM TL20MM
Common Device Name
SCREW, FIXATION, BONE
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ  2545
Manufacturer (Section G)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ   2545
Manufacturer Contact
sharon rivas
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key13281706
Report Number0008031020-2022-00020
Device Sequence Number12056468
Product Code HWC
UDI-Device Identifier04546540087638
UDI-Public04546540087638
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K000080
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number602640S
Device Catalogue Number602640S
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/21/2021
Supplement Date Received by Manufacturer03/17/2022
Initial Report FDA Received Date01/18/2022
Supplement Report FDA Received Date04/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-