• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL ASD, INC. MEDEX STOPCOCKS; STOPCOCK, I.V. SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD, INC. MEDEX STOPCOCKS; STOPCOCK, I.V. SET Back to Search Results
Model Number MX4341L
Medical Device Problem Code Obstruction of Flow (2423)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/01/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the stopcock primed and added to primary line.Free flows well but once hooked up to pump and pump programmed reads downstream occlusion and increased pressure.
 
Additional Manufacturer Narrative
This mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).No problems or issues were identified during this device history record review.A product sample was received for evaluation.Visual and functional testing were performed.One unopened sample was returned.The sample was not the sample that was reported as defected for occlusion.This was known due to the fact that the sample was returned in the sterile packaging unopened.Upon opening the sample the part was visually inspected under 10x magnification and found no signs of occlusion or intermittent occlusion in the part.Next, the sample was taken to the lab and water leak tested which includes priming each port of the sample.All testing came back positive with no occlusions or leaks.Next, the sample was tested per m.90 for plug/body hole alignment.This test requires that a 0.105 inch gauge pin be able to pass through all ports and plug holes with no greater than a 1/2 pound (lb) push.All holes were clear for the pin to pass through and required no pressure to enter.The root cause of the reported issue was found to be after review of the component, no signs of occlusion were found present.This sample was not the reported sample however so this does not mean that the sample(s) that were used as the customer facility did not experience this defect.If the defective components could be returned a more thorough investigation could be completed on the actual affected part.Based on the report, it is possible that the pump that is being used has a low backpressure alarm that could result from any interference in the line.Interferences could come in many forms.One such interference could be a diameter size change in the line due to the fluid traveling through the line having to change from one diameter size down to another smaller line, this would result in increased fluid velocity through the smaller line but may have a small back pressure effect from the wall section that is hit at the end of the larger diameter section.Recommendation was to check all lines and components for occlusion.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDEX STOPCOCKS
Common Device Name
STOPCOCK, I.V. SET
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6250 shier rings road
dublin OH 43016
Manufacturer (Section G)
NULL
6250 shier rings road
dublin OH 43016
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key13282312
Report Number3012307300-2022-00980
Device Sequence Number4861968
Product Code FMG
UDI-Device Identifier10351688507648
UDI-Public10351688507648
Combination Product (Y/N)N
Initial Reporter CountryCA
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/11/2024
Device Model NumberMX4341L
Device Catalogue NumberMX4341L
Device Lot Number4183215
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/07/2022
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/20/2021
Supplement Date Received by Manufacturer03/03/2023
Initial Report FDA Received Date01/18/2022
Supplement Report FDA Received Date04/12/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/15/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-