• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CARDINAL HEALTH NPWT FACILITY NPWT SVED THERAPY DEVICE; NEGATIVE PRESSURE WOUND THERAPY POWERED SUCTION PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CARDINAL HEALTH NPWT FACILITY NPWT SVED THERAPY DEVICE; NEGATIVE PRESSURE WOUND THERAPY POWERED SUCTION PUMP Back to Search Results
Model Number 6701132
Medical Device Problem Codes Overheating of Device (1437); Suction Problem (2170)
Health Effect - Clinical Codes Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/21/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The complaint was forwarded to the manufacturing facility where it is currently still under investigation.A follow-up report will be filed once the results have been completed.
 
Event or Problem Description
Customer reported that when the power was turned on the device, there was no suction even after troubleshooting.The device was not alarming to alert there was no suction.New device was obtained.There was no injury or adverse event reported.
 
Additional Manufacturer Narrative
Supplemental report being filed since sample device received for evaluation.Device released from qa to stock on 10/27/2021.Job router was reviewed without issue, service records were reviewed without issues.Serial number (b)(6) device passed all tests.From the sample investigation, it was discovered that the root cause was overheating which damaged tubing and wires causing pump failure.Capa-ma029-01812 was opened and an investigation into the overheating complaints was conducted.The following corrective actions were identified to address the root causes identified in the capa: these corrective actions are planned to be completed by december 2021.Ifu for sved device to be updated to include the warning for overheating and no direct contact of the device with patient.Update the service procedure to consider changing the motor is the motor has serviced more than 2000 hours.Implement the service timer recording in the service procedure and forms to determine when to change the motor once the serviceable life has been exhausted.Install relay cut off mechanism at appropriate threshold temperature that will cut off the current to the motor in cases when motor may draw excessive current leading to overheating effect.Cardinal health will continue to monitor the trend of this type of incident.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NPWT SVED THERAPY DEVICE
Common Device Name
NEGATIVE PRESSURE WOUND THERAPY POWERED SUCTION PUMP
Manufacturer (Section D)
CARDINAL HEALTH NPWT FACILITY
8322 eagle palm drive
riverview FL 33578
Manufacturer (Section G)
CARDINAL HEALTH NPWT FACILITY
8322 eagle palm drive
riverview FL 33578
Manufacturer Contact
patricia tucker
1500 waukegan rd
waukegan, IL 60085
8478874151
MDR Report Key13282499
Report Number1423537-2022-00717
Device Sequence Number5060404
Product Code OMP
UDI-Device Identifier10885380097454
UDI-Public10885380097454
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 03/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number6701132
Device Catalogue Number6701132
Device Lot Number14205
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/21/2021
Supplement Date Received by Manufacturer12/21/2021
Initial Report FDA Received Date01/18/2022
Supplement Report FDA Received Date03/20/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient Age69 YR
Patient SexFemale
Patient Weight49 KG
-
-