| Model Number |
94-0019 |
| Medical Device Problem Code |
Use of Device Problem (1670)
|
| Health Effect - Clinical Codes |
Pain (1994); Joint Laxity (4526)
|
| Date of Event |
12/07/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Clinical notification received for revision of left knee to address instability.Date of implantation: (b)(6) 2020.Date of revision: (b)(6) 2021.(left knee).Treatment: revision of insert component.Dor: (b)(6) 2021.Patient received a revision of the left knee insert to treat pain secondary to global instability.Upon entering the joint, a complete synovectomy was performed, and the instability was confirmed.After the insert was revised, the surgeon notes that there was still minimal varus/valgus instability secondary to a slightly internally rotated femoral component.The surgeon did not elect to revise the femoral component.The procedure was completed without complications.Doi: unk.Dor: (b)(6) 2021 left knee insert.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint #: (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.All available x-rays and photographs were reviewed, and it can't be confirmed any device related problem.Depuy synthese considers the investigation closed at this time.Should additional information be received, the information will be reviewed, and the investigation may be re-opened as necessary.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
|
| |
|
Search Alerts/Recalls
|