| Model Number |
G14834 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/29/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Occupation: other non-healthcare professional: materials coordinator.This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.
|
| |
|
Event or Problem Description
|
|
As reported, during an unspecified procedure a c-flex double pigtail ureteral stent set was found broken upon opening the packaging.It is unknown how the procedure was completed.No portion of the device was left within the patient.There was no need for additional procedures or prolonged hospitalizations.The patient did not experience any adverse effects due to this occurrence.Additional information has been requested.
|
| |
|
Event or Problem Description
|
|
Additional information was received on 17jan2022 but not included on the initial report: the tether was attached.The procedure was completed by using another stent.
|
| |
|
Additional Manufacturer Narrative
|
|
Blank fields on this form indicate the information is unknown, unavailable, or unchanged.The event is currently under investigation.A follow-up report will be submitted upon receipt of additional information or completion of the investigation.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
| |
|
Additional Manufacturer Narrative
|
|
Blank fields on this form indicate the information is unchanged, unknown or unavailable.Corrected information: b5 additional information: d4, h4 event description: as reported, during an unspecified procedure a c-flex double pigtail ureteral stent set was found broken upon opening the packaging (reference this report).Another c-flex multi-length ureteral stent set was opened and "shredded" while they were attempting to place it in the patient (reference mdr # 1820334-2022-00077 or mdr # 1820334-2022-00075.It is unknown which of these two stents "shredded" during this procedure).The procedure was completed by using another stent.No portion of the device was left within the patient.There was no need for additional procedures or prolonged hospitalizations.The patient did not experience any adverse effects due to this occurrence.Investigation ¿ evaluation: interview of personnel and a visual inspection of the returned device was conducted.A document based investigation was also performed including a review of complaint history, device history record (dhr), manufacturing instructions, specifications, the instructions for use (ifu), and quality control data.One c-flex double pigtail ureteral stent was returned for investigation in a ziploc bag with the label from the original packaging.One of the stent¿s pigtails was separated from the main stent body approximately 2.8 cm from the tip.The point of the separation had a stretched, almost burst-like appearance.A review of the device history record found no non-conformances related to the reported failure mode.A review of complaint history records shows no other complaints associated with the complaint device lot.Because there were no related non-conformances, adequate inspection activities have been established, there is objective evidence that the dhr was fully executed, and no other lot related complaints that have been received from the field, it was concluded that there is no evidence that nonconforming product exists in house or in the field.A review of relevant manufacturing and quality control documents was conducted and determined the stent shape was adequately inspected during quality control.Cook has concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.The instructions for use (ifu), provides the following information to the user related to the reported failure mode: precautions improper handling can seriously weaken the stent.¿ how supplied upon removal from the package, inspect the product to ensure no damage has occurred.¿ the cause of the stent separation was not able to be determined.Per the quality engineering risk assessment, no further action is required.The appropriate personnel have been notified, and we will continue to monitor for similar complaints.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
| |
|
Event or Problem Description
|
|
-additional information was received 17jan2022, but inadvertently not reported in our initial or follow up medwatch reports: there were two c-flex double pigtail ureteral stent sets used in the unspecified procedure.One of the stents was found broken upon opening the packaging captured in this report.The other stent "shredded" while they were trying to place it in the patient (reference mdr # 1820334-2022-00077 or mdr # 1820334-2022-00075).It is unknown which of these two stents "shredded" during this procedure.
|
| |
|